Applications 5.15
Measuring point identification 5.16
Fct. 3.06 APPLICAT.
Press key
.
FLOW = Set characterization for the flow, press key
STEADY (flow is steady)
PULSATING (pulsating flow, e.g. due to reciprocating pumps,
see also Sect. 6.4, 6.5 and 6.6
“Special-case applications”)
Press key to return to Fct. 3.06 APPLICAT.
}
select function using
key or
Fct. 3.07 MEAS.PNT
.
Press key
.
Set measuring point identification using max. 10 characters ,
(e.g. TQ1_532197)
You can use any out of A...Z / a...z / 0...9 or _ (underscore)
- Select character by pressing key or ↓.
- Use key to shift one place to the left, back with key.
Press key to return to Fct. 3.07 MEAS.PNT.
Communication interface 5.17
Control input E (IFC 020 E only) 5.18
Fct. 3.08 COM
Press key
.
Specify function
OFF (Switched off)
HART (HART Interface)
KROHNE (RS 485 Interface)
Use key or to set ADDRESS 00-15 after selecting "HART" with key.
Press key dto return to Fct. 3.08 COM.
Use key or to set ADDRESS 000-239 after selecting "KROHNE" (RS 485). Press key and
set up BAUD RATE: 1200 2400 4800 9600 19200. Use key or to select baud rate.
Press key to return to Fct. 3.08 COM.
}
select function using key or
Fct. 1.08 CTRL.INP
. E
Press key .
Select control input function, using key or .
OFF (Switched off, out of function)
ZERO OUTPUT (Set output and display to "Min. values")
RESET COUNT. (Reset counter)
RESET ERROR (Delete / confirm error message)
Press key to return to Fct. 1.08 CTRL.INP. E
For factory-set preferences refer to the start-up protocol and Section 3.2.