7/4
Group P Fault / Symptom Cause Remedial action
P 1 Totalizer connected but does Incorrect connection/polarity Connect properly,
not count any pulses see Sect. 2.4.
Totalizer or external voltage Check output (see Sect. 7.2)
source defective. with new totalizer:
T
est ok, check connection
cables and previous
totalizers and external voltage
source, and replace if
necessary.
T
est faulty, pulse output
defective, replace signal
converter (see Sect. 8.4)
or contact Krohne Service.
Current output is external Check connection and cables,
voltage source, short circuit see Sect. 2.4. Voltage between
or current / pulse output I
+
and Iapprox. 15 V.
defective Switch off device, eliminate
short circuit, switch device
on again. If no function,
then current or pulse
output defective. Replace
signal converter (see Sect.
8.4) or contact Krohne
Service.
Pulse output is deactivated, Switch on under Fct. 1.06.
see Fct. 1.06.
P 2 Unsteady pulse rate – Process product Increase or switch on
conductivity too low, time constant, under Fct. 1.02
particles/air inclusions orswitch on if necessary.
too large or inhomogeneous
– Pulsating flow
– Time constant too low
or switched off
P 3 Pulse rate too high or Incorrect setting for Change setting under Fct. 1.06.
too low. pulse output.
Group S Fault / Symptom Cause Remedial action
S 1 No function Incorrect connection/polarity Connect properly,
of status display see Sect. 2.4.
Status display or output Set status output under Fct. 1.07
defective or external to “F/R INDIC.“ (flow direction)
voltage source not and check (see Sect. 7.2)
supplying voltage. with new status display:
T
est ok, check previous
status display and external
voltage source, and replace
if necessary.
T
est faulty, status output
defective, replace signal
converter (see Sect. 8.4) or
contact Krohne Service.
Group D/I/P/S Fault / Symptom Cause Remedial action
D/I/P/S1 Unsteady display and – Process product Increase time constant,
outputs conductivity too low, under Fct. 1.02
particles/air inclusions
too large or inhomogeneous
– Pulsating flow
– Time constant too low
D/I/P/S 2 No display and no Power OFF Switch power on.
function of outputs
Check power fuse(s) Replace if defective,
F1 (F1 and F2 for DC). see Sect. 8.2.