4 ELECTRICAL CONNECTIONS
24
IFC 050
www.krohne.com 08/2013 - 4002184002 - MA IFC 050 R02 en
1 Strip the conductor to dimension a.
2 If a stranded drain wire is present, remove the shield that is present. Make sure not to damage
the stranded drain wire.
3 Slide an insulating tube over the stranded drain wire.
4 Crimp a wire end ferrule onto the stranded drain wire.
5 Crimp the wire end ferrules on the conductors.
6 Pull a shrinkable tube over the prepared cable.
Figure 4-4: Field current cable C, preparation for the signal converter
a = 80 mm / 3.15"