4 ELECTRICAL CONNECTIONS
30
IFC 050
www.krohne.com 08/2013 - 4002184002 - MA IFC 050 R02 en
To open the cover of the electrical terminal compartment, lightly press in the side walls of the
mains cover 2.
Flip the mains cover up.
Connect the power supply.
Close the mains cover again by flipping it down.
100...230 VAC (tolerance range: -15% / +10%)
100...230 VAC (tolerance range: -15% / +10%)100...230 VAC (tolerance range: -15% / +10%)
100...230 VAC (tolerance range: -15% / +10%)
Note the power supply voltage and frequency (50...60 Hz) on the nameplate.
24 VDC (tolerance range: -30% / +30%)
24 VDC (tolerance range: -30% / +30%)24 VDC (tolerance range: -30% / +30%)
24 VDC (tolerance range: -30% / +30%)
Note the data on the nameplate!
When connecting to functional extra-low voltages, provide a facility for protective separation
(PELV) (acc. to VDE 0100 / VDE 0106 and/or IEC 364 / IEC 536 or relevant national
regulations).
Figure 4-9: Terminal compartment for power
1 Cable entry for power supply
2 Mains cover
3 Ground terminal
4 100...230 VAC (-15% / +10%)
5 24 VDC (-30% /+30%)
INFORMATION!
240 VAC + 5% is included in the tolerance range.