Manual
ELECTRICAL CONNECTIONS 4
41
IFC 050
www.krohne.com08/2013 - 4002184002 - MA IFC 050 R02 en
Status output / limit switch passive
• U
ext
≤ 32 VDC
• I ≤ 100 mA
• R
L
= 47 kΩ
• closed:
U
0
≤ 0.2 V at I = 10 mA
U
0
≤ 2 V at I = 100 mA
open:
I ≤ 0.05 mA at U
ext
= 32 V
INFORMATION!
•
Independent of connection polarity.
•
The output is open if the signal converter is not powered.
Figure 4-17: Status output / limit switch passive S
p