Manual

6 OPERATION
52
IFC 050
www.krohne.com 08/2013 - 4002184002 - MA IFC 050 R02 en
6.4.2 Menu B, test
No. Function Setting / Description
B1 simulation
B1.1 volume flow
mass flow
Depending on the selection, simulation of the volume or mass flow
according to the following sequence:
Select:
set value /
break (exit function without simulation)
Query: start simulation?
Select:
no (exit function without simulation) / yes (start simulation)
B1.2 current output A simulation X
[X stands for one of the connection terminals A or D]
Sequence and settings similar to B1.2, see above!
For pulse output a set number of pulses are output in 1 s!
B1.3 pulse output D
B1.3 frequency out X
B1.3 limit switch D
B1.3 status output D
B2 actual values
B2.1 operating hours Display the actual operating hours; exit the displayed function with key ^.
B2.2 act. flow speed Display the actual flow speed; exit the displayed function with key ^.
B2.3 electr. temperature Display the actual electronics temperature; exit the displayed function with
key ^.
B2.4 act. coil resistance Display the actual resistance of the field coils.
B3 information
B3.1 C number This number clearly identifies the electronics.
B3.2 Electronic Revision ER Reference identification number, electronic revision and production date of
the device; includes all hardware and software changes
B3.3 device serial no. Serial no. of the system.
B3.4 electronic serial no. Serial no. of the electronics