6 OPERATION
64
IFC 050
www.krohne.com 08/2013 - 4002184002 - MA IFC 050 R02 en
6.5.2 Deleting error messages in the menu "quick setup"
6.6 Status messages and diagnostic information
Operational faults in device
INFORMATION!
For the detailed list of the possible error messages refer to Status messages and diagnostic
information on page 64
.
Key Display Description and setting
> quick setup Press and hold for 2.5 s, then release the key.
> language -
3 x ↓ reset -
> reset errors -
> reset?
no
-
↓ or ↑ reset?
yes
-
^ reset errors Error has been reset.
3 x ^ Measuring mode -
Messages on the display Description Actions
Status: F _ _ _ _ _ Operational fault in device, mA output ≤
3.6 mA or set fault current (depending
on the seriousness of the fault), status
output open, pulse / frequency output:
no pulses
Repair necessary.
F error in device Error or failure of device. Parameter or
hardware error. No measurement possible.
Group message, when one of the following
or some other severe error occurs.
F IO 1 Error, operational fault in IO 1. Parameter
or hardware error. No measurement
possible.
Load settings (Fct. C4.6.3) (backup 1,
backup 2 or factory settings). If status
message still does not disappear, replace
electronic unit.
F parameter Error, operational fault of data manager,
electronic unit, parameter or hardware
error. Parameters no longer usable.
F configuration Invalid configuration: display software, bus
parameter or main software do not match
existing configuration.
If device configuration unchanged:
defective, replace electronic unit.
F display Error, operational fault in display.
Parameter or hardware error. No
measurement possible.
Defective, replace electronic unit.
F sensor electronic Error, operational fault in sensor
electronics. Parameter or hardware error.
No measurement possible.
Defective, replace electronic unit.
F sensor global Data error in the global data of the
measuring sensor electronic equipment.
Load settings (Fct. C5.6.3) (backup 1,
backup 2 or factory settings). If status
message still does not disappear, replace
electronic unit.
F sensor local Data error in the local data of the measuring
sensor electronic equipment.
Defective, replace electronic unit.
F field current local Data error in the local data of the field
current supply.
Defective, replace electronic unit.