8.5 Measuring accuracy
Reference conditions
Medium: water
Temperature: +20°C / +68°F
Pressure: 1 bar / 14.5 psi
Inlet section: 5 DN
X [m/s]: flow velocity
Y [%]: deviation from the actual measured value (mv)
DN [mm] DN [inch] Accuracy Curve
OPTIFLUX 2050 251200 148 0.5% of measured value
above 0.5 m/s 1
1
OPTIFLUX 1050 10150 3/86
OPTIFLUX 6050 10150 3/86
WATERFLUX 3050 25...600 1...24
1 below 0.5 m/s, deviation of ±2.5 mm/s