Quickstart EN

ELECTRICAL CONNECTIONS 3
25
IFC 100
www.krohne.com07/2010 - 4000124104 - QS IFC 100 R04 en
Open the cover of the electrical terminal compartment by pressing down and pulling forwards
at the same time.
Close the cover after the power has been connected.
100...230 VAC (tolerance range: -15% / +10%)
100...230 VAC (tolerance range: -15% / +10%)100...230 VAC (tolerance range: -15% / +10%)
100...230 VAC (tolerance range: -15% / +10%)
Note the power supply voltage and frequency (50...60 Hz) on the nameplate.
12...24 VDC (tolerance range: -55% / +30%)
12...24 VDC (tolerance range: -55% / +30%)12...24 VDC (tolerance range: -55% / +30%)
12...24 VDC (tolerance range: -55% / +30%)
Note the data on the nameplate!
When connecting to functional extra-low voltages, provide a facility for protective separation
(PELV) (acc. to VDE 0100 / VDE 0106 and IEC 364 / IEC 536 or relevant national regulations).
24 VAC/DC (tolerance range: AC: -15% / +10%; DC: -25% / +30%)
24 VAC/DC (tolerance range: AC: -15% / +10%; DC: -25% / +30%)24 VAC/DC (tolerance range: AC: -15% / +10%; DC: -25% / +30%)
24 VAC/DC (tolerance range: AC: -15% / +10%; DC: -25% / +30%)
AC: Note the power supply voltage and frequency (50...60 Hz) on the nameplate.
DC: When connecting to functional extra-low voltages, provide a facility for protective
separation (PELV) (acc. to VDE 0100 / VDE 0106 and IEC 364 / IEC 536 or relevant national
regulations).
Figure 3-12: Power supply connection
1 100...230 VAC (-15% / +10%), 8 VA
2 24 VDC (-55% / +30%), 4 W
3 24 VAC/DC (AC: -15% / +10%; DC: -25% / +30%), 7 VA and 4 W
INFORMATION!
240 VAC + 5% is included in the tolerance range.
INFORMATION!
12 VDC - 10% is included in the tolerance range.
INFORMATION!
12 V is not
notnot
not included in the tolerance range.