3.4.10 Preparing field current cable C, connection to measuring sensor
The field current cable is not included in delivery.
The shield for field current cable C can be connected to the measuring sensor.
Bending radius: 50 mm / 2"
Required materials
Heat-shrinkable tubing
3 wire end ferrules to DIN 46 228: size according to the cable being used
1 Strip the conductor to dimension a.
2 Remove any shield that is present.
3 Pull a shrinkable tube over the prepared cable.
4 Crimp the wire end ferrules onto the conductors 7, 8 and 9.
INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.
Figure 3-12: Field current cable C, preparation for the measuring sensor
a = 50 mm / 2"