Quickstart EN Instruction Manual

ELECTRICAL CONNECTIONS 3
47
IFC 300
www.krohne.com08/2010 - 4000070303 - QS IFC 300 R06 en
3.6.9 Preparing field current cable C, connection to measuring sensor
The field current cable is not part of the scope of delivery.
The shield is connected in the terminal compartment of the converter directly via the shield
and a clip.
The shield is connected in the sensor via the special cable gland.
Bending radius: 50 mm / 2"
Required materials
Shielded 2-wire insulated copper cable
Insulating tube, size according to the cable being used
Heat-shrinkable tubing
DIN 46 228 wire end ferrules: size according to the cable being used
1 Strip the conductor to dimension a.
2 Trim the outer shield to dimension b and pull it over the outer sheath.
3 Crimp wire end ferrules onto both conductors.
INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.
Figure 3-30: Preparation of field current cable C
a = 125 mm / 5"
b = 10 mm / 0.4"