Quickstart EN Instruction Manual
2 INSTALLATION
6
IFC 300
www.krohne.com 08/2010 - 4000070303 - QS IFC 300 R06 en
Possible scope of delivery for signal converter / measuring sensor
2.3 Storage
• Store the device in a dry, dust-free location.
• Avoid continuous direct sunlight.
• Store the device in its original packing.
• Storage temperature: -50...+70°C / -58...+158°F
2.4 Transport
Signal converter
• No special requirements.
Compact version
• Do not lift the device by the signal converter housing.
• Do not use lifting chains.
• To transport flange devices, use lifting straps. Wrap these around both process connections.
2.5 Installation specifications
Measuring sensor Measuring sensor + signal converter IFC 300
Compact Remote field
housing
Remote wall-
mounted housing
Remote rack-mounted
housing
R (28 TE) or (21 TE)
OPTIFLUX 1000 OPTIFLUX 1300 C OPTIFLUX 1300 F OPTIFLUX 1300 W OPTIFLUX 1300 R
OPTIFLUX 2000 OPTIFLUX 2300 C OPTIFLUX 2300 F OPTIFLUX 2300 W OPTIFLUX 2300 R
OPTIFLUX 4000 OPTIFLUX 4300 C OPTIFLUX 4300 F OPTIFLUX 4300 W OPTIFLUX 4300 R
OPTIFLUX 5000 OPTIFLUX 5300 C OPTIFLUX 5300 F OPTIFLUX 5300 W OPTIFLUX 5300 R
OPTIFLUX 6000 OPTIFLUX 6300 C OPTIFLUX 6300 F OPTIFLUX 6300 W OPTIFLUX 6300 R
OPTIFLUX 7000 OPTIFLUX 7300 C - - -
WATERFLUX 3000 WATERFLUX 3300 C WATERFLUX 3300 F WATERFLUX 3300 W WATERFLUX 3300 R
TIDALFLUX 4000 - TIDALFLUX 4300 F - -
INFORMATION!
The following precautions must be taken to ensure reliable installation.
•
Make sure that there is adequate space to the sides.
•
Protect the signal converter from direct sunlight and install a sun shade if necessary.
•
Signal converters installed in control cabinets require adequate cooling, e.g. by fan or heat
exchanger.
•
Do not expose the signal converter to intense vibration. The flowmeters are tested for a
vibration level in accordance with IEC 68-2-3.