Quickstart EN Instruction Manual

ELECTRICAL CONNECTIONS 3
63
IFC 300
www.krohne.com08/2010 - 4000070303 - QS IFC 300 R06 en
3.10.4 19" rack-mounted housing (21 TE), electrical connection of the inputs and outputs
For frequencies above 100 Hz, shielded cables are to be used in order to reduce effects from
electrical interferences (EMC).
Terminal A+ is only operable in the basic version.
Connect the conductor to the multipolar plug according to the illustration.
Press the plug into the connector.
3.10.5 Laying electrical cables correctly
1 Lay the cable in a loop just before the housing.
2 Tighten the screw connection of the cable entry securely.
3 Never mount the housing with the cable entries facing upwards.
4 Seal cable entries that are not needed with a plug.
DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take
note of the voltage data on the nameplate!
Figure 3-42: Terminal compartment for inputs and outputs in rack-mounted housing
A+ A+ C C-A-A
DD- BB-
Figure 3-43: Protect housing from dust and water