Manual

7 TECHNICAL DATA
68
OPTIBAR DP 7060 C
www.krohne.com 04/2014 - 4003292001 - MA OPTIBAR DP 7060 C R01 en
Electrical connections
Mechanical
MechanicalMechanical
Mechanical
Cable gland M20 x 1.5 mm
Cable diameter 5...9 mm / 0.20...0.35"
6...12 mm / 0.24...0.47"
10...14 mm / 0.39...0.55"
Cable entry
1/2 NPT
Blind plug: M20 x 1.5 mm, 1/2-14 NPT
Closing cap: M20 x 1.5 mm, 1/2-14 NPT
Connector option: M12 x 1, Harting HAN 7D,8D, 7/8 FF
Core cross-section
Solid wire with cords: 0.2 mm...2.5 mm
2
/ AWG 24...14
Cord with wire end sleeve: 0.2 mm...1.5 mm
2
/ AWG 24...16
Electrical
ElectricalElectrical
Electrical
Supply voltage Non-Ex device: 9.6...35 VDC
Ex ia device: 9.6...30 VDC
Ex d device: 9.6...35 VDC
Ex ia d device: 15...35 VDC
Reverse polarity
protection
Integrated
Permissible residual ripple
Non-Ex devices for U
n
12 VDC (9.6 < UB < 14 VDC) 0.7 V
eff
(16...400 Hz)
Ex ia devices,
Ex ia d devices
for U
n
24 VDC (18 < UB < 35 VDC) 1.0 V
eff
(16...400 Hz)
Load R
L,max
=(UB-9.6) / 22 mA
Potential connection
in device
Electronics: not electrically isolated
Ground terminal: galvanically connected with process connection
Overvoltage category III
Protection class II
Inputs and outputs
Output signal
Output signalOutput signal
Output signal
Output signal
4...20 mA / HART
®
version 7.3
3.8...20.5 mA (factory setting acc. to. NAMUR recommendation)
Signal resolution 0.3 µA
Error signal of current
output (adjustable)
High alarm 21 mA
Low alarm 3.6 mA, last valid measurement
Max. output current 21.5 mA
Boot-up current 10 mA for 5 ms after switching on, then 3.6 mA
Damping 0...999 seconds, adjustable
.book Page 68 Friday, May 2, 2014 10:25 AM