INSTALLATION 3
25
OPTIFLEX 2200 C/F
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
The probe counterweight has a hole with an M8 internal thread. You can attach the probe
counterweight to a threaded end.
Weld a tube with an internal diameter of 23...25 mm / 0.91...1¨ to the bottom of the tank.
i Make sure the tube aligns with the process connection at the top of the tank.
Lower the probe into the tank.
Put the end of the probe into the tube.
If this is not possible, you can attach braces to the probe.
Single cable Ø2 mm / 0.08¨
Figure 3-16: How to attach a Ø2 mm / 0.08¨ single cable probe to keep it straight
Coaxial Ø22 mm / 0.8¨
Figure 3-17: How to attach a coaxial probe to keep it straight