Manual

INSTALLATION 3
37
OPTIFLEX 2200 C/F
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
Use a 2 mm Allen wrench to attach and tighten a HC M4×20 screw at the top of each rod
segment (intermediary and end rod segments)
Attach a PTFE spacer to the end of each rod segment that has a groove.
Attach a pair of lock washers at the top of each rod segment (intermediary and end rod
segments)
Assemble one of the middle rod segments (with a pair of lock washers on the attached screw)
and the signal rod below. Use two 7 mm open-ended wrenches to tighten the assembled parts
to a torque of 23Nm.
Figure 3-30: How to assemble the segmented coaxial probe: part 1
CAUTION!
Do not attach the screw to the end of the rod segment that has a groove for the attachment of a
PTFE spacer.