Quickstart EN Owner's manual

2 INSTALLATION
12
OPTIFLEX 2200 C/F
www.krohne.com 08/2013 - 4000622103 - QS OPTIFLEX 2200 R03 en
For double cable and double rod probes:
For coaxial probes:
If your device has a coaxial probe, you can ignore these installation recommendations.
2.7.3 Installation requirements for concrete roofs
Figure 2-12: Recommended nozzle dimensions for double rod and double cable probes
d 50 mm / 2¨, where d is the diameter of the tank nozzle
CAUTION!
Install coaxial probes in clean liquids that are not too viscous.
Figure 2-13: Installation on a concrete roof
1 The diameter, d, of the hole must be greater than the thickness, t, of the concrete.
2 If the thickness, t, of the concrete is greater than the diameter, d, of the hole, install the device in a recess.