2 INSTALLATION
22
OPTIFLEX 2200 C/F
www.krohne.com 08/2013 - 4000622103 - QS OPTIFLEX 2200 R03 en
Attach a lock nut to each end of the rod probe segments.
Attach a union nut to the bottom end of each rod probe segment, but not to the bottom
segment. Use two 8 mm open-end wrenches to tighten the union nut against the lock nut.
Attach the top segment of the rod probe to the union nut below the process connection. Use
two 8 mm open-end wrenches to tighten the union nut against the lock nut on the rod probe.
Attach the middle segment of the rod probe to the union nut on the top segment (if there are
middle segments). Use two 8 mm open-end wrenches to tighten the union nut against the lock
nut. Repeat this step for the other segments.
Attach the bottom segment of the rod probe to the union nut on the top segment. Use two
8 mm open-end wrenches to tighten the union nut against the lock nut.
Figure 2-24: How to assemble the segmented single rod probe: part 2
WARNING!
Put a support below the probe to prevent deformation.
CAUTION!
Make sure that the nuts are tight and the rod probe cannot loosen.