2 INSTALLATION
26
OPTIFLEX 2200 C/F
www.krohne.com 08/2013 - 4000622103 - QS OPTIFLEX 2200 R03 en
Use a 2.5 mm Allen wrench to attach and tighten a HC M5×5 screw (lock screw) to the bottom
segment of tube.
2.10.4 How to install a device with a flange connection
Equipment needed:
Device
Gasket (not supplied)
Wrench (not supplied)
Make sure the flange on the nozzle is level.
Make sure that you use the applicable gasket for the flange and the process.
Align the gasket correctly on the flange facing of the nozzle.
Lower the probe carefully into the tank.
i For more data on cable probes, refer to
How to install a cable probe in the tank
on page 28.
Tighten the flange bolts.
i Refer to local rules and regulations for the correct torque to apply to the bolts.
Figure 2-28: How to assemble the segmented coaxial probe: part 3
CAUTION!
If the lock screw is not tight, the device will not measure correctly.
Figure 2-29: Flange connection