Quickstart EN Owner's manual
OPERATION 4
47
OPTIFLEX 2200 C/F
www.krohne.com08/2013 - 4000622103 - QS OPTIFLEX 2200 R03 en
• [>
>>
>] to change Scale 4 mA.
• [>
>>
>] to change the position of the cursor.
• [
] to decrease the value or [
] to
increase the value.
• [^
^^
^] to confirm.
Use this step to give the 4 mA output
setting (0% limit) in the tank. Refer to the
illustrations that follow. Illustration 1
shows the settings for level. Illustration 2
shows the settings for distance.
• [>
>>
>] to change Scale 20 mA.
• [>
>>
>] to change the position of the cursor.
• [
] to decrease the value or [
] to
increase the value.
• [^
^^
^] to confirm.
Use this step to give the 20 mA output
setting (100% limit) in the tank. Refer to
the illustrations that follow. Illustration 1
shows the settings for level. Illustration 2
shows the settings for distance.
• [
] or [
] for the selection of the error
delay (0 s, 10 s, 20 s, 30 s, 1 mn, 2 mn,
5mn or 15mn).
• [^
^^
^] to confirm.
The time after which the current output
changes to an error value. The error value
shows that there is a measurement error.
• [>
>>
>] to change the tag name.
• [>
>>
>] to change the position of the cursor.
• [
] to decrease the alphanumeric value
(A, B, ..., 1, 2, ...) or [
] to increase the
alphanumeric value.
• [^
^^
^] to confirm.
• 2× [^
^^
^] to confirm.
• [
] or [
] for the selection of the save
option (STORE NO or STORE YES).
• [^
^^
^] to confirm.
Set to STORE YES to save and use the data.
Set to STORE NO to cancel the changes to
the device settings.
Screen Steps Description