User guide

INSTALLATION 2
15
OPTIFLEX 2200 C/F
www.krohne.com04/2013 - 4000861902 - AD IECEX OPTIFLEX2200 R02 en
Remote (Field) versions (probe housing only)
EPL Ga/Gb: Ex ia and Ex d ia devices
Remote (Field) version (probe housing only)
EPL Gb: Ex ia and Ex d ia devices
EPL Gc: Ex ic devices
Temperature
class
Maximum ambient temperature Max. flange
temperature
Ø2 mm / Ø0.08¨
single cable probe
(HT version)
Ø2 mm / Ø0.08¨
single cable probe
(other versions)
All other probe types
[°C] [°F] [°C] [°F] [°C] [°F] [°C] [°F]
T6 +51 +123 +49 +120 +49 +120 +60 +185
T5 +70 +158 +70 +158 +70 +158 +60 +185
T4 +80 +176 +80 +176 +80 +176 +60 +185
Temperature
class
Minimum ambient temperature Min. flange
temperature
Ø2 mm / Ø0.08¨
single cable probe
(HT version)
Ø2 mm / Ø0.08¨
single cable probe
(other versions)
All other probe types
[°C] [°F] [°C] [°F] [°C] [°F] [°C] [°F]
T6...T2 -40 -40 -40 -40 -40 -40 -20 -4
INFORMATION!
Remote converter
Remote converterRemote converter
Remote converter
The maximum ambient temperature, T
a
, permitted depends on the temperature class:
T
a
= +60
°
C for class T6
T
a
= +70
°
C for class T5
T
a
= +80
°
C for class T4
Temperature
class
Maximum ambient temperature Max. flange
temperature
Ø2 mm / Ø0.08¨
single cable probe
(HT version)
Ø2 mm / Ø0.08¨
single cable probe
(other versions)
All other probe types
[°C] [°F] [°C] [°F] [°C] [°F] [°C] [°F]
T6 +51 +123 +49 +120 +49 +120 +60 +140
+48 +118 +39 +102 +43 +109 +85 +185
T5 +66 +150 +64 +147 +64 +147 +75 +167
+65 +149 +54 +129 +58 +136 +100 +212
T4 +79 +174 +77 +170 +78 +176 +85 +185
+75 +167 +64 +147 +68 +154 +110 +230
+71 +160 +51 +124 +59 +138 +135 +275