User guide
2 INSTALLATION
16
OPTIFLEX 2200 C/F
www.krohne.com 04/2013 - 4000861902 - AD IECEX OPTIFLEX2200 R02 en
T3 +69 +156 +43 +109 +54 1 +129 1 +150 +302
+65 +149 - - - - +180 2 +356 2
+62 +143 - - - - +200 2 +392 2
T2 +54 +129 - - - - +250 2 +482 2
+47 +116 - - - - +300 2 +572 2
1 If the device has an adaptor for BM 102 probes, this temperature is only permitted for devices that have an extension option
2 Make sure that gasket temperature is in the specified limits. For more data, refer to the handbook.
Temperature
class
Minimum ambient temperature Min. flange
temperature
Ø2 mm / Ø0.08¨
single cable probe
(HT version)
Ø2 mm / Ø0.08¨
single cable probe
(other versions)
All other probe types
[°C] [°F] [°C] [°F] [°C] [°F] [°C] [°F]
T6...T2 -40 -40 -40 -40 -40 -40 -40 1 -40 1
-39 -38 -35 -31 -36 -32 -50 1 -58 1
1 Make sure that gasket temperature is in the specified limits. For more data, refer to the handbook.
INFORMATION!
Remote converter
Remote converterRemote converter
Remote converter
The maximum ambient temperature, T
a
, permitted depends on the temperature class:
•
T
a
= +60
°
C for class T6
•
T
a
= +70
°
C for class T5
•
T
a
= +80
°
C for class T4
Temperature
class
Maximum ambient temperature Max. flange
temperature
Ø2 mm / Ø0.08¨
single cable probe
(HT version)
Ø2 mm / Ø0.08¨
single cable probe
(other versions)
All other probe types
[°C] [°F] [°C] [°F] [°C] [°F] [°C] [°F]