CONTENTS 1 OPTIFLUX 2070 Safety instructions......................................................................... 4 1.1 Intended use......................................................................................................... 4 1.2 Safety instructions from the manufacturer......................................................... 4 1.2.1 Disclaimer.................................................................................................................. 4 1.2.
CONTENTS OPTIFLUX 2070 5.1.1 6 Internal battery ........................................................................................................ 21 Operation ..................................................................................... 22 6.1 Display and operating elements ........................................................................ 22 6.2 Menu of IFC 070 converter................................................................................. 22 7 Service ............
1 SAFETY INSTRUCTIONS OPTIFLUX 2070 1.1 Intended use The electromagnetic flowmeter OPTIFLUX 2070 has been designed for measuring potable water and water with suspended particles. The OPTIFLUX 2070 features not only accurate flow measurement, but also continuous diagnostics in accordance with applicable standards. This self-diagnosis monitors and automatically reports improper functioning of the electronics, or faulty sensor electrodes.
SAFETY INSTRUCTIONS 1 OPTIFLUX 2070 1.2.3 Display conventions The following symbols are used to help you navigate this documentation more easily: WARNING! These warning signs must be observed without fail. Even only partial disregarding such warnings can result in serious health damage, damage to the device itself or to parts of the operator ’s plant. DANGER! This symbol designates safety advice on handling electricity. CAUTION! These warnings must be observed without fail.
1 SAFETY INSTRUCTIONS OPTIFLUX 2070 1.3.1 Transportation, handling and using instruction for batteries WARNING! The used Lithium batteries are primary power sources with high energy content. If mistreated, they may present a potential risk. NOTE! KROHNE assumes no liability for customer failure. Please observe the following transportation, handling and using instructions: • Transport only in special packaging with special labels and transportation documents.
INSTRUMENT DESCRIPTION 2 OPTIFLUX 2070 2.1 Scope of delivery • Check the packing list to check if you received all that you ordered. • Inspect the cartons carefully for damage or signs of rough handling. Report damage to the carrier and to your local KROHNE office. Figure 2-1: Scope of delivery 1 2 3 4 5 6 OPTIFLUX 2070 with blue protection cap Quick Start Factory calibration report CD-ROM including Handbook, Quick Start, Technical datasheet Grounding ring (optional) Cable (remote version only) 2.
2 INSTRUMENT DESCRIPTION OPTIFLUX 2070 2.3 Nameplate Figure 2-3: Nameplate OPTIFLUX 2070 1 2 3 4 5 6 Manufacturer Voltage information Material of wetted parts Meter constant Serial number Device type CAUTION! Check on the device nameplate. that the device is according to your order. 8 www.krohne.
INSTALLATION 3 OPTIFLUX 2070 3.1 Pre-installation requirements Make sure that you have all necessary tools available: • • • • • Allen key (4 mm) Philips screwdriver Wrench for cable glands Wrench for wall mounting bracket (remote version only) Torque wrench for installing flowmeter in pipeline 3.2 General installation notes GK070: 10.234 Figure 3-1: Visual check 3.3 Storage • Store the flowmeter in its original packing • storage temperature: -40 ... +65 °C 3.
3 INSTALLATION OPTIFLUX 2070 3.5 Installation requirements 3.5.1 Inlet and outlet Figure 3-3: Recommended inlet and outlet 1 ≥ 5 DN 2 ≥ 2 DN 3.5.2 Mounting position and flange deviation Figure 3-4: Mounting position and flange deviation 1 Lmax 2 Lmin • Mount flowmeter either with converter aligned upwards or downwards. • Install flowmeter in line with the pipe axis. • Pipe flange faces must be parallel to each other. 10 www.krohne.
INSTALLATION 3 OPTIFLUX 2070 CAUTION! Max. permissible deviation of pipe flange faces: Lmax - Lmin ≤ 0.5 mm 3.5.3 Vibration Figure 3-5: Vibration 3.5.4 Magnetic field Figure 3-6: Magnetic field 3.5.5 Bends Figure 3-7: Installation in bending pipes 04/2008 • 4000243601 www.krohne.
3 INSTALLATION OPTIFLUX 2070 3.5.6 Open discharge Install meter on a lowered section of the pipe to ensure a full pipe condition through the meter. Figure 3-8: Preferable installation close to open discharge 3.5.7 T-section Figure 3-9: Recommended inlet for T-section installation 1 ≥ 10 DN 12 www.krohne.
INSTALLATION 3 OPTIFLUX 2070 3.5.8 Control valve Always install control valves downstream of flowmeter in order to avoid cavitation or distortion of flow profile. Figure 3-10: Position of control valve 3.5.9 Pump Never install flowmeter at a pump suction side in order to avoid cavitation or flashing into the flowmeter. Figure 3-11: Preferable position of pump 04/2008 • 4000243601 www.krohne.
3 INSTALLATION OPTIFLUX 2070 3.5.10 Grounding Figure 3-12: The grounding of the flowmeter 1 Metal pipelines, not internally coated. Grounding without grounding rings! 2 Metal pipelines with internal coating and non-conductive pipelines. Grounding with grounding rings! 3 Grounding rings CAUTION! The grounding of the flowmeter provides a stable and accurate measurement. 3.6 Installation Here you find the maximum pressure and torques for the flowmeter.
INSTALLATION 3 OPTIFLUX 2070 Nominal size [inch] Flange class [lbs] Bolts Max. torque [Nm] Polypropylene Hardrubber 2 150 4 × 5/8" 24 23 3 150 4 × 5/8" 43 39 4 150 8 × 5/8" 34 31 5 150 8 × 3/4" 48 43 6 150 8 × 3/4" 61 51 8 150 8 × 3/4" - 69 10 150 12 x 7/8" - 79 12 150 12 x 7/8" - 104 14 150 12 x 1" - 93 16 150 16 x 1" - 91 18 150 16 x 1 1/8" - 143 20 150 20 x 1 1/8" - 127 24 150 20 x 1 1/4" - 180 Procedure to setup max.
3 INSTALLATION OPTIFLUX 2070 3.7 Mounting of converter NOTE! Only applicable for remote versions 3.7.1 Mounting of IFC 070 F • Mount IFC 070 F with mounting plate on wall or standpipe. • Keep distance between sensor and signal converter as short as possible. • Observe length of the delivered signal cable. 16 www.krohne.
ELECTRICAL CONNECTIONS 4 OPTIFLUX 2070 4.1 Safety instructions WARNING! Observe the regional occupational health and safety regulations without fail. Only work on the device electrics if you are appropriately trained. DANGER! Observe national installation regulations! 4.2 Terminal assignment of IFC 070 Figure 4-1: Removing side cap 04/2008 • 4000243601 www.krohne.
4 ELECTRICAL CONNECTIONS OPTIFLUX 2070 Figure 4-2: Terminal assignment 1 2 3 4 5 6 Status output 1 Status output 2 Not connected Ground Pulse output A Pulse output B Electrical values • Pulse output passive: f ≤ 500 Hz; I ≤ 10 mA; U: 5...24 VDC (P ≤ 100 mW) • Status output passive: I ≤ 10 mA; U: 5...24 VDC (P ≤ 100 mW) 4.3 Signal cable CAUTION! Always use the supplied KROHNE signal cable. NOTE! You only receive a signal cable if you ordered a remote version. 18 www.krohne.
ELECTRICAL CONNECTIONS 4 OPTIFLUX 2070 4.4 Cable connection for field version 4.4.1 Flow sensor side Figure 4-3: Cable connection at flow sensor side 1 2 3 4 5 6 cable length: 13 cm / 5" cable length: 5 cm / 2" cable length: 8 cm / 3" brown + white cable, used for field current purple + blue cable, used for electrode signals shield (terminal 1 of connector X2 + U-clamp) • Prepare appropriate cable lengths (1...3).
4 ELECTRICAL CONNECTIONS OPTIFLUX 2070 4.4.2 Converter side Figure 4-4: Cable connection at converter side 1 2 3 4 5 6 cable length: 16 cm / 6.3" cable length: 5 cm / 2" cable length: 10 cm / 4" brown + white cable, used for field current purple + blue cable, used for electrode signals shield (terminal 1 of connector X2 + U-clamp) • Prepare appropriate cable lengths (1...3). • Connect the shield to the U-clamp, the brown cable to terminal 7 and the white to terminal 8.
START-UP 5 OPTIFLUX 2070 5.1 Connection of battery CAUTION! Please connect battery before first use. KROHNE delivers each IFC 070 converter with a disconnected battery. 5.1.
6 OPERATION OPTIFLUX 2070 6.1 Display and operating elements 1 2 3 4 5 6 battery status optical key V to navigate through the menu and to scroll through the measuring pages reset button (only accessible with removed cover) flow direction measured value and measuring unit optical key Z to navigate through the menu 6.
OPERATION 6 OPTIFLUX 2070 CAUTION! Take care with changing menu number 13. If you set this to "1", the display locks. IF this happens: • Remove the blue protection cap • Remove the 4 Allen keys (4 mm) • Remove the cover • Press both keys together with the reset button for 6 seconds as shown in the graph below • Use a small screwdriver for the reset button • Display starts with the menu, beginning with menu no. 12 • Go to menu no.
6 OPERATION OPTIFLUX 2070 Software version 2.0.0 Menu No. Description Display default Selection list Remarks 12 Measuring unit x 0 = m3, 1 = USG Customer defined 13 Fiscal metering 0 0 = No, 1 = Yes Blocks menu 21 Meter size xxx Defined at factory callibration - 22 Meter constant xx.xxx Defined at factory callibration - 23 Zero offset calibration 0 Set to 1 In situ determination of zero point. Confirm with “>” for 3 seconds After countdown meter switches to measuring mode.
OPERATION 6 OPTIFLUX 2070 Software version 2.0.0 Menu No.
7 SERVICE OPTIFLUX 2070 7.1 Replacement of battery Figure 7-1: Removing battery • Remove the blue protection cap. • Remove the 4 Allen keys (4 mm). • Remove the cover. • Disconnect the connector of the battery. • Remove the battery holder by pulling it upwards. Figure 7-2: Connecting battery • Remove the battery from the holder. • Insert the new battery in the holder. • Replace the holder. • Fasten the battery connector to the internal connector in the IFC 070 converter.
SERVICE 7 OPTIFLUX 2070 • Replace the cover. WARNING! Make sure that the battery cable is not jammed by the cover. • Tighten the 4 bolts. • Enter the programming mode, hold the V and Z key for 5 seconds. i The display flashes. • Press the Z key to enter the menu. • Scroll through the available positions with the V key to position 74 and check the battery capacity (important for battery life indication). • Go to position 75 and enter "1" (reset battery counter).
8 TECHNICAL DATA OPTIFLUX 2070 8.1 Technical data OPTIFLUX 2070 flowmeter Versions Compact OPTIFLUX 2070 C: IFC 070 C converter on top of OPTIFLUX 2000 flow sensor Field OPTIFLUX 2070 F: IFC 070 F converter separate from OPTIFLUX 2000 flow sensor Performance Measurement functionality Default: totalised volume Selectable: actual volume flowrate, + counter totaliser, - counter totaliser Measuring range 0…12m/s (0…39ft/s) Max. deviation (under reference conditions) 1 ±0.
TECHNICAL DATA 8 OPTIFLUX 2070 600 24" 500 20" 450 18" 400 16" 350 14" 300 12" 250 10" 200 8" 150 6" 125 5" 3" 80 100 4" 2 1/2" 65 DN [mm] VN15 2" ASME [inch] VN14 50 Nominal diameter Liner Polypropylene Hardrubber See pressure and temperature limits for various liners Electrodes Hastelloy C4 Stainless steel 1,4571 (AISI 316 Ti) Titanium Grounding rings Hastelloy C4 Stainless steel 1,4571 (AISI 316 Ti) Titanium Flanges Steel 1.0460 (C 22,8 ) Steel 1.
8 TECHNICAL DATA 600 24" 500 20" 450 18" 400 16" 350 14" 300 12" 250 10" 200 8" 150 6" 125 5" 3" 80 100 4" 2 1/2" 65 DN [mm] VN15 2" ASME [inch] VN14 50 Nominal diameter OPTIFLUX 2070 Versions Compact Separate Electrical conductivity Min. conductivity 50 µS/cm J standard J optional U on request Maximum deviation Y [%] vs flow velocity X [m/s] 1 OPTIFLUX 2070 30 www.krohne.
TECHNICAL DATA 8 OPTIFLUX 2070 IFC 070 flow converter Functions Flow direction Bi-directional measurement Low flow cut-off Selectable Design Display 8 digits LCD Status indication for battery, diagnostics Units m3, US Gallons m3/h, USGPM Cable connections Standard 2x M20 x 1,5 Optional 1/2" NPT, PF 1/2 Materials Converter housing Die-cast aluminium (polyurethane coated) In- and output Outputs 2 passive pulse outputs for remote totalising f ≤ 500 Hz; I ≤ 10 mA; U: 5...
8 TECHNICAL DATA OPTIFLUX 2070 Temperature range Process [°C] min. Ambient [°C] max. min. max. Hardrubber Separate flow sensor (OPTIFLUX 2000 F) -5 80 -40 65 Compact with IFC 070 (OPTIFLUX 2070 C) -5 80 -40 65 Separate flow sensor (OPTIFLUX 2000 F) -5 90 -40 65 Compact with IFC 070 (OPTIFLUX 2070 C) -5 90 -40 65 Polypropylene 1 1 Polypropylene available for DN 50 - 150 Process [°F] min. Ambient [°F] max. min. max.
TECHNICAL DATA 8 OPTIFLUX 2070 8.2 Dimensions and Weights OPTIFLUX 2070 C T 070 H W OPTIFLUX 2000 F c a Tbox W IFC 070 F 310mm 335mm 175mm 04/2008 • 4000243601 www.krohne.
8 TECHNICAL DATA OPTIFLUX 2070 Dimensions and weights in mm and kg Flanges acc. EN 1092-1 DN PN Dimensions [mm] L H T box W T 070 Approx.
TECHNICAL DATA 8 OPTIFLUX 2070 Dimensions and weights in mm, bar and kg Flanges acc. ASME B 16.5 DN 04/2008 • 4000243601 PN Approx.
OPTIFLUX 2070 nnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn KROHNE Product Overview • • • • • • Electromagnetic flowmeters Variable area flowmeters Mass flowmeters Ultrasonic flowmeters Vortex flowmeters • • • • • Level measuring instruments Pressure gauges Temperature measuring instruments Water solutions & analysis Oil and gas turnkey solutions Flow controllers Addresses: Germany KROHNE sales companies Northern sales office KROHNE Messtechnik GmbH & Co. KG Bremer Str.