Manual

Installation and Operating Instructions OPTIMASS
19
1.2.9 Technical Data
Nominal Flow Rates
06 10 15 25 40 50 80
Kg/h
950 2,700 11,250 34,500 91,500 180,000 430,000
Lbs/min
35 100 400 1,250 3,350 6,600 15,800
Maximum flow rate
Typically 130 % of the nominal flow rate for the sensor size depending on application.
Minimum flow rate
Depending on measuring error required.
Tube materials:
Titanium Gr. 9,
Hastelloy C22 and
SS 318.
The meter size has a prefix T, H, or S indicating the tube material.
Secondary pressure containment
All 70 Series meters have secondary containment rated to 40 bar or 580 psi.
An optional 63 bar or 914 psi is available.
Materials of construction
Flanges: SS 316 L
Spigots and outer cylinder: SS 304 L optional SS 316 L
Front end housing and post: SS 316 L
Converter housing: Epoxy coated aluminium