Installation and Operating Instructions OPTIMASS
23
1.3 OPTIMASS 3000 (7100) Single Z Shaped Tube Meter
1.3.1 Specific Installation Guidelines
When installing, please observe the following:
• Four holes are provided in the base plate and all four should always be used.
• The plastic inserts in the base plate mounting holes are important to ensure a rigid and stable
connection to the mounting structure.
• It is important to mount on a firm and rigid structure to obtain a stable zero condition.
• The following guidelines are provided to assist the installer to select the best option:
Flow
Vertical mount is possible. Horizontal mount is possible.
.
Flow
Take care to avoid gas build up in top half of tube
bends.
Do not suspend from flanges alone. Base must
always be supported.
Do not install upside down