2 Electrical Installation
2.1 Location and connecting cables
Location
Do not expose the compact flow meter to direct sunlight in hot climates. Install a sunshade if necessary.
Connecting cables
To conform to protection category requirements, observe the following recommendations:
Fit blanking plug and apply sealant to unused cable entries.
Do not kink cables directly at cable entries.
Provide water drip point (U bend in cable).
Do not connect rigid conduit to cable entries.
Only cables of diameter 7 to 12 mm or ¼” to ½” can be used.
2.2 Connection to power
Please ensure that the information about power given on the data plate corresponds to the locally
available mains voltage.
Note information given on the instrument data plate (voltage, frequency)!
Electrical connection in conformity with IEC 364 or equivalent national standard. Special regulations
apply to installation in hazardous areas. (See supplementary installation and operating instructions)
The PE protective ground conductor must be connected to the separate U-clamp terminal in the
terminal box of the signal converter.
Do not cross or loop the cables in the terminal box of the signal converter. Use separate cable glands
for power and output cables.
Ensure that the screw thread of the round cover on the terminal box is well greased at all times.
Note:
The grease used must be non-corrosive to aluminium; typically it must be resin- and acid-free.
Protect sealing ring from damage.
2.2.1 Power Supply Wiring MFC050
Power and signal connections for MFC 050
2.2.2 Power Supply Wiring MFC051 Non Ex
24 V AC/DC 100 – 230 V AC mains
Outputs and
connections
PE
N L AC
– + DC
L N
PE
Outputs
24 V AC/DC
PE