4.4 Table of programmable functions
Fct.
No
Text Description and settings
1 OPERATION Main menu 1 Operation
1.1 ZERO CALIB. Submenu 1.1 Zero adjustment
1.1.1 AUTO. CALIB. Automatic zero adjustment
*1) Select: SURE YES or NO
*2) If YES: Calibration (approx. 30 sec. duration)
Display: Actual flow rate as percent of the maximum rated flow for the primary
head. (Q
100%
)
*3) Select: ACCEPT YES or NO
1.1.2 MANUAL CAL. Input manual zero offset
* Direct input of a zero flow offset.
Units: As selected by Fct.
1.1.3 DISP. ZERO Display of the last zero in percent of nom. flow
1.2 INST. STATE Input of the instrument state
Use the key to switch between three states of operation, then press :
*MEASURE
*STANDBY (tube vibrating, Mass Flow set to zero)
*STOP (tube drive stopped)
1.3 DENSITY.CAL Submenu 1.3 Density calibration
1.3.1 DISP. PT. 1 Display of last density calibration value point 1
1.3.2 DISP. PT. 2 Display of last density calibration value point 2
1.3.3 1 POINT.CAL. Density calibration mode: 1 point calibration
* SURE (NO). Use the key to select YES, then press
Use the key to select desired calibration sample from the list below:
*
EMPTY
* WATER
* TOWN WATER
* OTHER
2 POINT.CAL. Density calibration mode: 2 point calibration
1
st
access to menu 1.3.4:
* SURE (NO). Use the key to select YES, then press . Use the key to select
between
* CAL.SAMPLE1
* EXIT
Press and use the key to select desired calibration sample from the list below then
press
* EMPTY
* WATER
* TOWN WATER
* OTHER
1.3.4
CALIB. OK.
Press to return to Fct. 1.3.4
2 POINT.CAL 2
nd
access to menu 1.3.4:
* SURE (NO). Use the key to select YES, then press . Use the key to select
between
* CAL.SAMPLE2
* RESTART
* EXIT
Press and use the key to select desired calibration sample from the list below then
press .
* WATER
* TOWN WATER
* OTHER
1.3.4
CALIB. OK
Press to return to Fct. 1.3.4
1.3.5 FACTORY.SET Reset factory settings
Restore density calibration to factory settings
* SURE (NO). Use the key to select YES, then press .