4.4 Table of programmable functions
Fct.
No
Text Description and settings
1 OPERATION Main menu 1 Operation
1.1 ZERO CALIB. Submenu 1.1 Zero adjustment
1.1.1 AUTO. CALIB. Automatic zero adjustment
*1) Select: SURE YES or NO
*2) If YES: Calibration (approx. 30 sec. duration)
Display: Actual flow rate as percent of the maximum rated flow for the primary
head. (Q
100%
)
*3) Select: ACCEPT YES or NO
1.1.2 MANUAL CAL. Input manual zero offset
* Direct input of a zero flow offset.
Units: As selected by Fct.
1.1.3 DISP. ZERO Display of the last zero in percent of nom. flow
1.2 INST. STATE Input of the instrument state
Use the key ↑ to switch between three states of operation, then press ↵:
*MEASURE
*STANDBY (tube vibrating, Mass Flow set to zero)
*STOP (tube drive stopped)
1.3 DENSITY.CAL Submenu 1.3 Density calibration
1.3.1 DISP. PT. 1 Display of last density calibration value point 1
1.3.2 DISP. PT. 2 Display of last density calibration value point 2
1.3.3 1 POINT.CAL. Density calibration mode: 1 point calibration
* SURE (NO). Use the ↑ key to select YES, then press ↵
Use the ↑ key to select desired calibration sample from the list below:
*
EMPTY
* WATER
* TOWN WATER
* OTHER
2 POINT.CAL. Density calibration mode: 2 point calibration
1
st
access to menu 1.3.4:
* SURE (NO). Use the ↑ key to select YES, then press ↵. Use the ↑ key to select
between
* CAL.SAMPLE1
* EXIT
Press ↵ and use the ↑ key to select desired calibration sample from the list below then
press ↵
* EMPTY
* WATER
* TOWN WATER
* OTHER
1.3.4
CALIB. OK.
Press ↵ to return to Fct. 1.3.4
2 POINT.CAL 2
nd
access to menu 1.3.4:
* SURE (NO). Use the ↑ key to select YES, then press ↵. Use the ↑ key to select
between
* CAL.SAMPLE2
* RESTART
* EXIT
Press ↵ and use the ↑ key to select desired calibration sample from the list below then
press ↵.
* WATER
* TOWN WATER
* OTHER
1.3.4
CALIB. OK
Press ↵ to return to Fct. 1.3.4
1.3.5 FACTORY.SET Reset factory settings
Restore density calibration to factory settings
* SURE (NO). Use the ↑ key to select YES, then press ↵.