4.5 Reset / Quit Menu - Totalizer reset and status indication acknowledgement
Totalizer reset
Button Display Description
10.36
kg
Measurement mode
CodE 2
––
Enter access Code 2 for reset/quit menu:
RESET.TOTAL Totalizer reset menu
If the additional totaliser is selected (Fct. 3.1.6) a choice of reset
options is presented:
* RESET ALL Reset all totalisers
* ADDITIONAL Reset only the additional totaliser
Otherwise the following is presented:
* SURE YES
* SURE NO
Note: The reset option can be disabled by Fct. 3.5.3 or Custody
Transfer (CT) lock.
View status message(s) and quit
Button Display Description
0.36
kg/min
Measurement mode
The presence of the marker above Status on the display
indicates the presence of warning messages in the status list.
CodeE 2
– –
Enter access code for reset/quit menu:
↑→
RESET.TOTAL
Totalizer reset menu.
STATUS.LIST
View/Quit Status message menu
MASS FLOW
Use either the or keys to view other messages in the list.
Otherwise press to exit.
QUIT YES
At the end of the message list the QUIT YES prompt is shown.
Selecting YES will clear if possible messages in the list.
To cancel the operation press to get QUIT NO and then press
.
STATUS.LIST Assuming the conditions that caused the message have passed
(i.e. mass flow is back within the meter’s range) then the Status
marker, will disappear.
Returns to measuring mode.
An overview with typical status messages and descriptions will be given in the table in chapter 6.2.
View FE Status
Button Display Description
CodeE 2
– –
Enter access code for reset/quit menu:
RESET.TOTAL
STATUS.LIST
FE STATUS
Totalizer reset menu.
View/Quit Status message menu
View FE Status messages
Messages
Normally no messages are present.
Further messages are sometimes present which indicate the
various diagnostic indicators, primarily for service or trouble