Viewing sensor signal conditions – Diagnose Fct. 2.9
Menu 2.9 allows the viewing of eight parameters
Display Key
Line 1 Line 2
Fct. 2.9 Diagnose
Fct. 2.9.1
This displays the temperature of the measuring tube.
By pressing the key, the temperature will be indicated.
By pressing the key, you will be returned to the Function display.
Fct. 2.9.2 Strain M. T.
This displays the strain value of the measuring tube stain gauge in ohms (where fitted).
Fct. 2.9.3 Strain I. C.
This displays the strain value of the inner cylinder strain gauge in ohms (where fitted).
Fct. 2.9.4 Tube Freq.
This displays the current resonant frequency of the sensor. This value is primarily used to
calculate the density of the process fluid.
Fct. 2.9.5 Drive energy
This value indicates the current to the driver in percent. The heavier the fluid, the higher the
number. Air entrainment also shows up as a higher number.
Fct. 2.9.6 Sensor A
Fct. 2.9.7 Sensor B
These display the sensor signal level. Should be as per Menu 5.2.4, and be within 2% of
each other.
Fct. 2.9.8 Comm.Errors
This displays the number of serial communication errors between the Front End (sensor
mounted) and the converter electronics since power up. Normally displays 0
Viewing hardware and software versions fitted Fct. 2.10
Menu 2.10 allows the viewing of the hardware and software versions installed in the mass flowmeter.
Display
Key
Line 1 Line 2
Fct. 2.10 Versions
Fct. 2.10.1 Backend SW
This displays the software version in the converter MFC 050/051
Fct. 2.10.2 Backend HW
This displays the hardware version of the converter MFC 050/051
Fct. 2.10.3 Frontend SW
This displays the software version installed in the front end electronics mounted on the
sensor.