Manual

INSTALLATION 3
11
OPTISENS CL 1100
www.krohne.com01/2012 - 4001834501 - MA OPTISENS CL 1100 R01 en
Storing the sensor in the provided plastic tube
Screw the sealing cap off the plastic tube. Keep it in the original packing.
If there is not enough KCl solution in the plastic tube, fill it up with 3 molar KCl solution.
Insert the sensor tip through the hole in the storage cap (refer to drawing on page 11).
Carefully push the O-ring delivered with the storage cap on the sensor so that the cap sits over
the O-ring.
Insert the sensor tip into the plastic tube until it is fully covered with KCI solution.
Tighten the cap.
Store the sensor in its original packing.
3.3 Pre-installation requirements
Unpacking the electrode
Loosen the storage cap which is screwed on to the plastic tube 1.
Gently pull the sensor out of the plastic tube 2.
Lay the sensor on a soft mat/tissue 3.
Screw the provided sealing cap on to the plastic tube, using O-ring and washer as pictured in
the drawing 4. Keep the storage cap (the one with the hole in it) in the original packing.
CAUTION!
Do not touch or scratch the gold electrodes of the sensor.
Make sure that the gold electrodes are clean and dust-free. If necessary, clean the tip as
described on page 35
.
Figure 3-2: Handling the electrode