Manual

OPERATION 5
23
OPTISENS CL 1100
www.krohne.com01/2012 - 4001834501 - MA OPTISENS CL 1100 R01 en
C1.13.4 clean Start cleaning procedure, manually.
C1.15
C1.15C1.15
C1.15
C2.15
C2.15C2.15
C2.15
temperature Menu for temperature measurement. Available for sensor 1 and sensor 2.
C1.15.1
C2.15.1
probe Options:
manual: used if no internal or external temperature sensor is connected to
the signal converter
Pt100: used if the pH electrode has an integrated Pt temperature
measurement or if an external Pt100 temperature sensor is connected to
the signal converter
Pt1000: used if an external Pt1000 temperature sensor is connected to the
signal converter
C1.15.2
C2.15.2
manual Only available if C1.15.1 or C2.15.1 is set to "manual".
C1.15.3
C2.15.3
correction Offset correction for temperature measurement. Not available if C1.15.1 or
C2.15.1 is set to "manual".
C1.15.4
C2.15.4
limitation Measuring ranges for temperature measurement.
C1.15.5
C2.15.5
temp. comp. Menu for activating the temperature compensation parameters.
For detailed information refer to
Temperature compensation
on page 24.
Options:
linear: linear temperature compensation.
off: temperature compensation is disabled.
C1.18
C1.18C1.18
C1.18 product calibration For detailed information refer to
Calibration
on page 24.
C1.18.1 temp.comp Menu for activating the temperature compensation parameters for the
calibration.
Options:
off: temperature measurement is disabled
manual: temperature value has to be entered manually
automatic: temperature measurement is performed as configured
C1.18.2 temperature Menu for setup of manual temperature measurement
C1.18.3 temp. coefficient Temperature coefficient for manual temperature compensation during
calibration.
C1.18.4 pH compensation Options:
off: pH measurement is disabled
manual: pH value has to be entered manually
automatic: pH measurement is performed as configured
C1.18.5 man. pH compensation If in menu C1.18.4 the pH compensation was set to manual, here the desired
pH value is entered manually.
C1.18.6 start calibration Start calibration procedure.
C1.18.7 stored value View stored value of calibration.
C1.18.8 reference Enter the reference value of the probe in mg/L.
CAUTION!
If you choose for measurement the temperature compensation "linear" than choose between
"automatic" or "manual" for the temperature compensation during calibration. If you choose for
measurement the temperature compensation "off" than choose also "off" for the temperature
compensation during calibration.
Level Designation / function Settings / descriptions
.book Page 23 Wednesday, February 1, 2012 11:58 AM