Completing the calibration procedure
As soon as the laboratory value has been determined, go again into the calibration menu and
submenu reference
referencereference
reference by using
or
and then press >
>>
>.
Enter the laboratory value using
or
and press ^
^^
^.
i The signal converter now matches the "stored value" to the laboratory value and makes a
re-calibration. The slope appears on the display and it is stored in the calibration logbook
(this helps to compare the slope values over the time when re-calibrating).
Assure that the slope of a new sensor is 50...150 mg/nA, otherwise the sensor has a
malfunction. Press ^
^^
^ if the slope is okay.
i The signal converter now asks store value?
store value?store value?
store value?
Using
or
choose yes
yesyes
yes to store the calibration values or no
nono
no to discard them.
Press ^
^^
^ to confirm.
Press ^
^^
^ several times until the questions appears if the configuration should be stored.
Using
or
choose yes
yesyes
yes to store the configuration or no
nono
no to discard it.
i You have completed the sensor calibration.
Step 5: switching back to measurement
Deactivate the function "manual hold" again.
INFORMATION!
If an error occurs during the calibration procedure, the signal converter displays an error
message. Possible causes for an error:
Slope too flat.
Sensor too old.
Wrong electrical connection.