Instruction Manual

INSTALLATION 3
15
OPTISENS PH 9100
www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en
The following instructions describe the connection of the different sensor cables.
Connecting the sensor cable to the signal converter
Thread the sensor cable through the outer right cable gland 1.
Push the coax shield 4 into terminal N 2 and the coax core into terminal O 3.
To remove a cable, press down the white clip 5 on the corresponding terminal and pull the
cable out 6.
3.5.2 Connecting the external temperature sensor
Connect an external Pt100 or Pt1000 sensor to terminal block Pos.A/B of the signal converter
according to the following drawings:
Figure 3-4: Connecting the sensor cable
Figure 3-5: Connection of an external Pt100/1000 temperature sensor to the signal converter
1 2-wire connection
2 3-wire connection