Instruction Manual

OPERATION 4
35
OPTISENS PH 9100
www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en
Press ^ to confirm.
i The slope of the sensor is displayed (should be near 59 mV in an intact sensor).
Press ^ to confirm.
i The message store cal. value? is displayed on the screen. The converter asks if the new
calibration values should be stored.
Choose yes to store the calibration values. Choose no to discard the results.
Press ^ to confirm.
i Prior returning to the measuring display, you are asked if the configuration should be
stored.
Choose yes using or to store the new calibration values.
i You have completed the pH calibration.
If you want to return to the measuring mode, press ^ several times until you reach this mode.
Step 5: re-installing the sensor
After the calibration procedure, rinse the sensor with tap water.
Reinstall the sensor into its assembly, refer to
Installing the sensor
on page 17.
Step 6: switching back to measurement
Deactivate the function "manual hold" again.
INFORMATION!
If an error occurs during the calibration procedure, the display shows an error message.
Possible causes for an error are:
Slope too flat.
Wrong buffer solution.
Wrong electrical connection.
Moisture in the sensor connector.