User guide

4 ELECTRICAL CONNECTIONS
28
OPTISONIC 7300 / 8300
www.krohne.com 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02
INFORMATION!
For converters with an
Ex e
terminal compartment, the compartment can be opened in an
energized state for short periods of time, to access the intrinsically safe terminals for possible
checks. However, the semi-circular insulation cover over the non-intrinsically safe mains supply
terminals L and N (or L+ and L-) MUST be kept closed.
Overview of possible "Ex ia" in-/outputs, defined by characters XYZ of the CG 36 numbers
Characters
XYZ
Name of I/O
circuits
Terminals
A, A-, A+
Terminals
B, B-
Terminals
C, C-
Terminals
D, D-
200 Exi-IO n.c. n.c. CO(a) PO/SO
300 n.c. n.c. CO PO/SO
210 Exi-IO with Exi-
Option
CO(a) PO/SO/CI CO(a) PO/SO
220 CO PO/SO/CI CO(a) PO/SO
310 CO(a) PO/SO/CI CO PO/SO
320 CO PO/SO/CI CO PO/SO
230 Exi-IO with Exi-
Option 2
Iin
a
active PO/SO/CI CO(a) PO/SO
330 Iin
a
active PO/SO/CI CO PO/SO
240 Iin
p
passive PO/SO/CI CO(a) PO/SO
340 Iin
p
passive PO/SO/CI CO PO/SO
D00 Fieldbus I/O
Profibus PA
n.c. n.c. PA PA
D10 Fieldbus I/O
Profibus PA
with Ex i Option
CO(a) PO/SO/CI PA PA
D20 CO PO/SO/CI PA PA
E00 Fieldbus I/O
Foundation
Fieldbus
n.c. n.c. FF FF
E10 Fieldbus I/O
Foundation
Fieldbus with
Ex i Option
CO(a) PO/SO/CI FF FF
E20 CO PO/SO/CI FF FF
Used abbreviations for in-/output functions: CO = Current Output, PO = Pulse Output, SO = Status Output,
CI = Control Input, n.c. = not connected
All in-outputs are passive unless otherwise noted as active with extension (a).