Operating Instructions Indicating/adjustment module for OPTISOUND sensors
Contents Contents 1 About this document 1.1 Contents . . . . . . . . . . . . . . . . . . . . . . . . . 1.2 Target group . . . . . . . . . . . . . . . . . . . . . . 1.3 Symbolism used. . . . . . . . . . . . . . . . . . . . 4 4 4 2 For 2.1 2.2 2.3 2.4 2.5 2.6 2.7 6 6 6 6 6 7 7 your safety Authorised personnel . . . . . . . . . . . Appropriate use . . . . . . . . . . . . . . . Warning about misuse . . . . . . . . . . General safety instructions . . . . . . . CE conformity . . . . . . . . . . . . . . . .
Contents 30603-EN-050401 8 Supplement 8.1 Technical data . . . . . . . . . . . . . . . . . . . . . 28 8.2 Dimensions . . . . . . . . . . . . . . . . . . . . . . .
About this document 1 About this document 1.1 Contents This operating instructions manual has all the information you need for quick setup and safe operation of the indicating/adjustment module. Please read this manual before you start setup. 1.2 Target group This operating instructions manual is directed to trained personnel. The contents of this manual should be made available to these personnel and put into practice by them. 1.
About this document Sequence Numbers set in front indicate successive steps in a procedure.
For your safety 2 For your safety 2.1 Authorised personnel All operations described in this operating instructions manual must be carried out only by trained and specialist personnel authorised by the operator. For safety and warranty reasons, any internal work on the instruments must be carried out only by personnel authorised by the manufacturer. 2.2 Appropriate use The indicating and adjustment module is a pluggable unit for OPTISOUND level sensors. 2.
For your safety Conformity has been judged acc. to the following standards: l l EMC: - Emission EN 61326: 1997 - Susceptibility EN 61326: 1997 + A1:1998 NSR: EN 61010-1: 2001 2.6 Compatibility acc. to NAMUR NE 53 The indicating/adjustment module meets NAMUR recommendation NE 53. The parameter adjustment of the basic sensor functions is independent of the software version. The available functions depend on the appropriate software version of the single components. 2.
Product description 3 Product description 3.1 Configuration Scope of delivery The scope of delivery encompasses: l l Configuration Indicating/adjustment module Documentation - this operating instructions manual. The indicating/adjustment module consists of a display with full dot matrix as well as four keys for adjustment. 1 2 Fig.
Product description 1 2 Fig. 2: Rear of the indicating/adjustment module 1 2 integrated seal ring gold-plated contact path 3.2 Principle of operation Area of application The indicating/adjustment module is used for measured value indication, adjustment and diagnosis for the following OPTISOUND sensors: OPTISOUND 3010 C OPTISOUND 3020 C l OPTISOUND 3030 C l OPTISOUND 3040 C l OPTISOUND 3050 C The indicating/adjustment module is integrated into the respective sensor housing.
Product description 3.3 Adjustment Adjustment is carried out via the integrated keys. The entered parameters are generally saved in the respective sensor. With a copy function, parameters can be loaded into the indicating/adjustment module. 3.4 Storage and transport Packaging Your instrument was protected by packaging during transport. Its capacity to handle normal loads during transport is assured by a test acc. to DIN EN 24180.
Mounting 4 Mounting 4.1 Mounting procedure Mounting/dismounting the indicating/adjustment module The indicating/adjustment module can be mounted or dismounted at any time. An interruption of the power supply is not necessary. Proceed as follows: 1 Unscrew the housing cover 2 Place the indicating/adjustment module to the requested position on the electronics Information: Four different positions are possible, each displaced by 90°. 30603-EN-050401 Fig.
Mounting Note: If you intend to retrofit OPTISOUND with an indicating/ adjustment module for continuous measured value indication, a higher cover with an inspection glass is required. Dismounting is carried out in reverse order.
Set up 5 Set up 5.1 Adjustment system 2 1 1.1 3 Fig.
Set up keys are shown in the above illustration. Approx. 10 minutes after the last pressing of a key, an automatic reset to measured value indication is triggered. Any values not confirmed with [OK] will not be saved. 5.2 General functions Introduction OPTISOUND ultrasonic sensors have various functions. Therefore they can be adapted perfectly to the respective application. These functions are structured in menu form. Some of the functions are sensor-specific.
Set up Damping To damp process-dependent measured value fluctuations, you have to set an integration time of 0 … 999 s in this menu item. Damping Linearization curve In this menu item you select the linearization curve: linear Cylindrical tank l Spherical tank l User programmable User programmable means: Switching on a linearization curve programmed via PC and PACTware™ l l The linearization curve relates height and volume.
Set up Peak values Meas. reliability When using sensors working acc. to the non-contact measuring principle, the measurement can be influenced by the respective process conditions. In this menu item, the reliability of the level echo is displayed in dB. The reliability is signal strength minus noise. The higher the value, the more reliable the measurement. Sensor status Indication "OK" or flashing error message, e.g. "E013". In addition, the error appears in clear text in the measured value display. Meas.
Set up "X-Zoom": Zoom function for the meas.
Set up Simulation Simulation start? With the reset function, modified values are reset. Three subfunctions are available: Reset l l l Basic adjustment - Reset of the values modified with the indicating/ adjustment module to the sensor-specific basic setting (see chart) Factory setting - Like basic setting, but also reset of special parameters modified with PACTware™ to delivery status Peak values measured value and temperature - Reset of the min./max.
Set up Reset Reset select? Units of measurement In this menu item you select the internal calculating unit of the sensor: m(d) or ft(d). Units of measurement m(d) Language The sensor is already set to the ordered national language. In this menu item you can change the language. The following languages are available: l l l l l Deutsch English Français Espanõl Pycckuu. Language Deutsch Copy sensor data With this function data are read from the sensor data are written into the sensor.
Set up Sensor data copy? PIN In this menu item, the PIN is activated/deactivated permanently. Entering a 4-digit PIN protects the sensor data against unauthorized access and unintentional modifications. If the PIN is activated permanently, it can be deactivated temporarily (i.e. for approx. 60 min.) in any menu item. The instrument is delivered with the PIN set to 0000.
Set up Sensor type Serial number Date of manufacture Software revision Date of last change using PC Sensor details 5.3 Special functions – 4 … 20 mA/HART Introduction The 4 … 20 mA/HART special functions are briefly described in this paragraph. The respective range of functions of the indicating and adjustment module is determined by the sensor and the sensor software revision. Display In the menu item "Display" you can define how the measured value should be presented on the display.
Set up Mass Flow l Volume l Without unit. Depending on the selection, the different units will be available. l l In the menu item "Scaling", the requested numerical value with decimal point is entered for 0 % and 100 % of the measured value. There is the following relation between the indication value in the menu "Display" and the adjustment unit in the menu "Basic adjustment": l With ultrasonics, the indication value "Distance" means: Presentation of the measured value in the selected adjustment unit, e.
Set up Current output Output mode 4 … 20 mA 20 … 4 mA Failure mode1) Hold value 20.5 mA 22.0 mA <3.6 mA Min. current2) 3.8 mA 4 mA The values in bolt font represent the data of the factory setting. In HART multidrop mode, the current is constantly 4 mA. This value does not change even in case of failure. Current output Output mode 4-20 mA Failure mode: 22mA Min-current 3.8mA HART mode HART offers standard and multidrop mode3). In multidrop mode up to 15 sensors can be operated on one two-wire cable.
Set up 5.4 Menu schematic Basic adjustment > Basic adjustment Display Diagnostics Service Info 1 Min. adjustment 0.00 % = 4.000 m(d) 3.000 m(d) 1.1 Damping 1.5 Max. adjustment 100.00 % = 1.000 m(d) 2.000 m(d) 1.2 Linearization curve 1.6 0s Medium Liquid 1.3 Vessel form 1.4 Storage tank Sensor-TAG linear 1.7 Sensor Display > Basic adjustment Display Diagnostics Service Info Displayed value 2 2.1 scaled Display units 2.2 Volume m³ Scaling 2.3 0% = 0.
Set up Service > Basic adjustment Display Diagnostics Service Info False echo storage 4 4.1 Addl. adjustments 4.2 none 4.5 Units of measurement Reset select? Copy sensor data 4.6 m(d) select? 4.9 Copy sensor data? 4.3 Output mode 4-20 mA Failure mode <3.6 mA min. current3.8 mA Change? Reset Current output PIN Language Simulation 4.4 Simulation start? 4.7 Deutsch HART mode 4.8 standard Address 0 4.10 Enable? Info > Basic adjustment Display Diagnostics Service Info 5.
Maintenance and fault rectification 6 Maintenance and fault rectification 6.1 Maintenance When used appropriately under normal condition, the indicating and adjustment module is maintenance-free. 6.2 Instrument repair If it is necessary to repair, please proceed as follows: You can download a return form from our Internet homepage http://www.krohne-mar.com/fileadmin/ media-lounge/PDF-Download/Specimen_e.pdf.
Dismounting 7 Dismounting 7.1 Dismounting procedure Warning: Before dismounting, be aware of dangerous process conditions such as e.g. pressure in the vessel, high temperatures, corrosive or toxic products etc. Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order. 7.2 Disposal The indicating and adjustment module consists of materials which can be recycled by specialised recycling companies.
Supplement 8 Supplement 8.
8.2 Dimensions Indicating/adjustment module 27,6mm (1 3/32") ø66,3mm (2 39/64") 45,1mm (1 25/32") 9,7mm (3/8") 30603-EN-050401 Fig.
603-EN-050401 30 Indicating/adjustment module for OPTISOUND sensors
30603-EN-050401 Indicating/adjustment module for OPTISOUND sensors 31