Instruction Manual
9 Supplement
9.1 Technical data
General data
Materials, wetted parts
- Process fitting Thread G1½A and 1½ NPT: PVDF
- Transducer PVDF
- Seal transducer/process fitting EPDM
Materials, non-wetted parts
- Housing plastic PBT (Polyester), Alu-die casting pow-
der-coated, 316L (1.4435)
- Seal ring between housing and
housing cover
NBR (stainless steel housing), silicone (Alu/
plastic housing)
- Inspection window in housing cover Polycarbonate
- Ground terminal 316Ti/316L (1.4571/1.
4435)
Weight 1.8 … 4.0 kg (4.0 … 8.8 lbs), depending on
process fitting and housing
Max. torque mounting boss 25 Nm
Output variable
Output signal 4 … 20 mA/HART
Resolution 1.6 µA
Fault signal current output unchanged; 20.5 mA; 22 mA;
<3.6 mA (adjustable)
Current limitation 22 mA
Load see load diagram in Power supply
Integration time (63 % of the input
variable)
0 … 999 s, adjustable
Fulfilled NAMUR
recommendation NE 43
Input variable
Parameter distance between lower edge of the transducer
and product surface
Dead zone 0.25 m (0.8 ft)
Measuring range
- Liquids up to 5 m (16.4 ft)
- Solid up to 2 m (6.6 ft)
38 OPTISOUND 3010 C - 4 … 20 mA/HART
Supplement
30503-EN-050914