Owner's manual

Modes (adjustable)
- AMax. detection or overow/overll protection
- BMin. detection or dry run protection
Measuring accuracy
Hysteresis approx. 2 mm (0.08 in) with vertical installation
Integration time approx. 500 ms
Frequency approx. 1200 Hz
Ambient conditions
Ambient temperature on the housing -40 +70 °C (-40 +158 °F)
Storage and transport temperature -40 +80 °C (-40 +176 °F)
Process conditions
Parameter Limit level of liquids
Process pressure -1 64 bar (-14.5 938
psi) depending on
the process tting, e.g. ange
OPTISWITCH 5100 C, 5150 C of 316L/
Hastelloy C4 (2.4610)
-50 +150 °C (-58 +302 °F)
Process temperature (thread or ange
temperature) with temperature adapter
(option)
- OPTISWITCH 5100 C, 5150 C of
316L/Hastelloy C4
-50 +250 °C (-58 +482 °F)
- OPTISWITCH 5100 C, 5150 C
enamelled
-50 +
200 °C (-58 +392 °F)
- OPTISWITCH 5100 C, 5150 C with
ECTFE coating
-50 +150 °C (-58 +302 °F)
- OPTISWITCH 5100 C, 5150 C with
PFA coating
-50 +150 °C (-58 +302 °F)
26 OPTISWITCH 5100 C, 5150 C - with relay output
Supplement
30426-EN-060831