.book Page 2 Wednesday, April 30, 2014 1:39 PM : IMPRINT ::::::::::::::::::::::::::::::::::::::: All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without the prior written authorisation of KROHNE Messtechnik GmbH. Subject to change without notice. Copyright 2014 by KROHNE Messtechnik GmbH - Ludwig-Krohne-Str. 5 - 47058 Duisburg (Germany) 2 www.krohne.
.book Page 3 Wednesday, April 30, 2014 1:39 PM CONTENTS OPTISYS SLM 2100 1 Safety instructions 1.1 1.2 1.3 1.4 5 Software history ............................................................................................................... 5 Intended use ..................................................................................................................... 5 Certifications ...................................................................................................................
.book Page 4 Wednesday, April 30, 2014 1:39 PM CONTENTS OPTISYS SLM 2100 5 Operation 5.1 5.2 5.3 5.4 36 Start-up and general remarks for configuration........................................................... 36 Switching on the power .................................................................................................. 37 Operating elements........................................................................................................ 38 Measuring page ............................
Safety instructions .book Page 5 Wednesday, April 30, 2014 1:39 PM SAFETY INSTRUCTIONS 1 OPTISYS SLM 2100 1.1 Software history INFORMATION! This documentation explains the installation and operation of the OPTISYS SLM 2100 sludge level meter. Release date Software version Documentation 04/2013 01.00.01 MA OPTISYS SLM 2100 R02 en 1.2 Intended use The OPTISYS SLM 2100 sludge level meter is primarily designed for use in water and waste water treatment plants.
.book Page 6 Wednesday, April 30, 2014 1:39 PM 1 SAFETY INSTRUCTIONS OPTISYS SLM 2100 1.
.book Page 7 Wednesday, April 30, 2014 1:39 PM SAFETY INSTRUCTIONS 1 OPTISYS SLM 2100 1.4 Safety instructions from the manufacturer 1.4.1 Copyright and data protection The contents of this document have been created with great care. Nevertheless, we provide no guarantee that the contents are correct, complete or up-to-date. The contents and works in this document are subject to copyright. Contributions from third parties are identified as such.
.book Page 8 Wednesday, April 30, 2014 1:39 PM 1 SAFETY INSTRUCTIONS OPTISYS SLM 2100 1.4.3 Product liability and warranty The operator shall bear responsibility for the suitability of the device for the specific purpose. The manufacturer accepts no liability for the consequences of misuse by the operator. Improper installation and operation of the devices (systems) will cause the warranty to be void.
.book Page 9 Wednesday, April 30, 2014 1:39 PM SAFETY INSTRUCTIONS 1 OPTISYS SLM 2100 1.4.5 Warnings and symbols used Safety warnings are indicated by the following symbols. DANGER! This warning refers to the immediate danger when working with electricity. DANGER! This warning refers to the immediate danger of burns caused by heat or hot surfaces. DANGER! This warning refers to the immediate danger when using this device in a hazardous atmosphere. DANGER! These warnings must be observed without fail.
Device description .book Page 10 Wednesday, April 30, 2014 1:39 PM 2 DEVICE DESCRIPTION OPTISYS SLM 2100 2.1 Scope of delivery INFORMATION! Inspect the packaging carefully for damages or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer. INFORMATION! Do a check of the packing list to make sure that you have all the elements given in the order. INFORMATION! Look at the device nameplate to ensure that the device is delivered according to your order.
.book Page 11 Wednesday, April 30, 2014 1:39 PM DEVICE DESCRIPTION 2 OPTISYS SLM 2100 2.
.book Page 12 Wednesday, April 30, 2014 1:39 PM 2 DEVICE DESCRIPTION OPTISYS SLM 2100 Cable drum assembly and heater unit The cable drum helps to move the sensor up and down in the basin. It is mounted directly on the motor shaft of a synchronous motor, hidden behind the drum mounting plate. The cable drum always moves with a very constant speed. Since the diameter of the cable roll in the drum varies, the speed of the sensor varies. That is why the sensor moves slower during the end of the sampling.
.book Page 13 Wednesday, April 30, 2014 1:39 PM DEVICE DESCRIPTION 2 OPTISYS SLM 2100 2.3 Nameplate INFORMATION! Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate. Figure 2-3: Example of a nameplate 1 2 3 4 5 6 7 8 9 Manufacturer CE marking Manufacturing date Power supply data Observe the operation and installation instruction Protection category Order code / TAG no.
Installation .book Page 14 Wednesday, April 30, 2014 1:39 PM 3 INSTALLATION OPTISYS SLM 2100 3.1 General notes on installation INFORMATION! Inspect the packaging carefully for damages or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer. INFORMATION! Do a check of the packing list to make sure that you have all the elements given in the order. INFORMATION! Look at the device nameplate to ensure that the device is delivered according to your order.
.book Page 15 Wednesday, April 30, 2014 1:39 PM INSTALLATION 3 OPTISYS SLM 2100 3.3 Installation order DANGER! Do not install the sludge level meter in hazardous areas, it can ignite explosive gases! DANGER! Do not cover or obstruct the ventilation. It can lead to overheating of the device. INFORMATION! Protect the device against strong weather (lightning protection / overvoltage protection) and sunlight to prevent damages.
.book Page 16 Wednesday, April 30, 2014 1:39 PM 3 INSTALLATION OPTISYS SLM 2100 3.3.1 Mounting of the sludge level meter INFORMATION! All needed bolts, nuts and washers belong to the standard scope of delivery. To ensure proper assembly, please use only the mounting material provided with the meter. INFORMATION! All blots, nuts and washers should be greased to prevent cold welding and ensure simple disassemble after use. INFORMATION! Please ask a second person to help with this procedure.
.book Page 17 Wednesday, April 30, 2014 1:39 PM INSTALLATION 3 OPTISYS SLM 2100 Step 2a: Mounting on round handrail • Mount the frame with help of the brackets in a position that the horizontal bar of the frame is above the handrail. Figure 3-2: Mounting on handrail Step 2b: Mounting on square handrail • Mount the frame with help of the brackets in a position that the horizontal bar of the frame is above the handrail.
.book Page 18 Wednesday, April 30, 2014 1:39 PM 3 INSTALLATION OPTISYS SLM 2100 Step 4: Adjusting • Open the door of the meter for 45°. • Adjust the meter horizontally using the last two screws that you screw in the lower part of the frame. • Turn the screws until the door does not move (open and close) any more from the 45° position. Figure 3-4: Adjusting frame 3.
.book Page 19 Wednesday, April 30, 2014 1:39 PM INSTALLATION 3 OPTISYS SLM 2100 3.5 Installing the cleaning unit (optional) The cleaning unit consist of a water connection, an OPTISYS SLM 2100 controlled valve and the spraying unit to clean the cable and the sensor. If installing a device containing a cleaning unit, keep a safety area around the sensor garage free from electrical device or water sensitive parts, as outlined in the following drawing. Figure 3-6: Safety area 1 Radius: 2 m / 78.
.book Page 20 Wednesday, April 30, 2014 1:39 PM 3 INSTALLATION OPTISYS SLM 2100 3.5.1 Installation of the water hose WARNING! The maximum allowed water pressure should not exceed 6 bar / 87 psi. INFORMATION! An external separation switch of the water supply have to be located close to the device and easily accessible. A 3/4" connector wit metric thread provides the mounting point for the water hose adapter. When installing the water hose adapter on the connector, carefully fix the nut of the fitting.
Electrical connections .book Page 21 Wednesday, April 30, 2014 1:39 PM ELECTRICAL CONNECTIONS 4 OPTISYS SLM 2100 4.1 Safety instructions DANGER! All work on the electrical connections may only be carried out with the power disconnected. Take note of the voltage data on the nameplate! DANGER! Observe the national regulations for electrical installations! WARNING! Observe without fail the local occupational health and safety regulations.
.book Page 22 Wednesday, April 30, 2014 1:39 PM 4 ELECTRICAL CONNECTIONS OPTISYS SLM 2100 4.3 Description of electrical symbols Symbol Description mA meter, 0...20 mA or 4...20 mA and other, RL is the internal resistance of the measuring point including the cable resistance Internal DC voltage source Controlled internal current source in the device Switch, NO contact or similar Input, galvanically insulated PTC resistance Symbol for LED Symbol for the protection switch 22 www.krohne.
.book Page 23 Wednesday, April 30, 2014 1:39 PM OPTISYS SLM 2100 ELECTRICAL CONNECTIONS 4 4.4 Important device-specific notes on electrical connection DANGER! Never install or operate the device in potentially explosive areas, it might cause an explosion that can result in fatal injuries! DANGER! The device must be grounded in accordance with regulations in order to protect personnel against electric shocks.
.book Page 24 Wednesday, April 30, 2014 1:39 PM 4 ELECTRICAL CONNECTIONS OPTISYS SLM 2100 4.5 Overview cable connections DANGER! When installing and wiring the device, note the safety regulations of the current state of the art. Also note the following items to avoid fatal injuries, destruction or damage of the device or measuring errors: • Do not run cables on pathways. Regularly check the cable runs that no loose cables are hanging into footpaths.
.book Page 25 Wednesday, April 30, 2014 1:39 PM ELECTRICAL CONNECTIONS 4 OPTISYS SLM 2100 4.6 Overview of the terminal compartment The terminal compartment is located in the electronic compartment at the lower part of the main board.
.book Page 26 Wednesday, April 30, 2014 1:39 PM 4 ELECTRICAL CONNECTIONS OPTISYS SLM 2100 4.7 Connecting the power supply DANGER! The device must be grounded in accordance with regulations in order to protect personnel against electric shocks. DANGER! Never install or operate the device in potentially explosive areas, it might cause an explosion that can result in fatal injuries! CAUTION! When connecting the power supply, always note the safety regulations of the current state of the art.
.book Page 27 Wednesday, April 30, 2014 1:39 PM OPTISYS SLM 2100 ELECTRICAL CONNECTIONS 4 INFORMATION! All cables must have a test voltage of min. 2 kV and an appropriate outer insulation (additional to the insulation of the individual wire). The outer insulation should be removed 40 mm / 1.57". The minimum cross section of the wires is 1.5 mm2 / 0.0023 square inch and the maximum is 2.5 mm2 / 0.0039 square inch.
.book Page 28 Wednesday, April 30, 2014 1:39 PM 4 ELECTRICAL CONNECTIONS OPTISYS SLM 2100 4.8.2 Control input (active) INFORMATION! The factory default is a disabled control input! The electrical properties of the control inputs in detail are the following: • Active • Umax ≤ 32 VDC Function of the control inputs The three active control inputs can trigger different events in the meter from outside. They are engaged via simple contact closure of an external contact.
.book Page 29 Wednesday, April 30, 2014 1:39 PM OPTISYS SLM 2100 ELECTRICAL CONNECTIONS 4 Maintenance mode The function of the maintenance mode is to set the device to maintenance mode in order to enable easy cleaning of sensor, cable and meter. For this purpose a NO switch has to be connected to CI 3 (Note the switch is not part of the scope of delivery. You have to buy this switch with another manufacturer).
.book Page 30 Wednesday, April 30, 2014 1:39 PM 4 ELECTRICAL CONNECTIONS OPTISYS SLM 2100 4.9.2 Current output (active) DANGER! All work on the electrical connections may only be carried out with the power disconnected. CAUTION! To avoid damage or destruction of the device always note the following items: • Observe the connection polarity! • Note the properties of the current output; for further information refer to Technical data on page 68.
.book Page 31 Wednesday, April 30, 2014 1:39 PM ELECTRICAL CONNECTIONS 4 OPTISYS SLM 2100 INFORMATION! All cables must have a test voltage of min. 2 kV and an appropriate outer insulation (additional to the insulation of the individual wire) . The outer insulation should be removed 30-50 mm / 1.18"1.97". The minimum cross section of the wires is 0.5 mm2 / 0.00077 square inch and the maximum is 1.5 mm2 / 0.0023 square inch.
.book Page 32 Wednesday, April 30, 2014 1:39 PM 4 ELECTRICAL CONNECTIONS OPTISYS SLM 2100 Figure 4-6: Control input 1 Signal The terminals for the connection of the three control inputs are located on the main board. Please refer to the following diagram for proper connection of the cables. Figure 4-7: Connection control input 1 Control input (CI1) 2 Control input (CI2) 3 Control input (CI3) Connecting the control inputs • Open the cable drum and electronic compartment door.
.book Page 33 Wednesday, April 30, 2014 1:39 PM OPTISYS SLM 2100 ELECTRICAL CONNECTIONS 4 4.10 Description and properties of the relays DANGER! Never install or operate the device in potentially explosive areas, it might cause an explosion that can result in fatal injuries. The two relays can be either configured as limit switch or as status output.
.book Page 34 Wednesday, April 30, 2014 1:39 PM 4 ELECTRICAL CONNECTIONS OPTISYS SLM 2100 Note the electrical properties of the relays. For more information refer to Technical data on page 68. The terminals for the connection of the two current relays are located on the main board. Please refer to the following diagram for proper connection of the cables.
.book Page 35 Wednesday, April 30, 2014 1:39 PM OPTISYS SLM 2100 ELECTRICAL CONNECTIONS 4 4.11 Electrical connection of the external pump DANGER! Never install or operate the device in potentially explosive areas, it might cause an explosion that can result in fatal injuries. DANGER! Never connect an external water pump directly to the OPTISYS SLM 2100 connector K3, the output connector is only specified as a control output for a motor protection.
Operation .book Page 36 Wednesday, April 30, 2014 1:39 PM 5 OPERATION OPTISYS SLM 2100 5.1 Start-up and general remarks for configuration DANGER! In order to avoid injuries or damage of the device or plant equipment the device shall only be operated with both doors properly closed and locked. The device also has to be properly mounted and all cable feedthrough has to be properly tightened or plugged, if unused.
.book Page 37 Wednesday, April 30, 2014 1:39 PM OPERATION 5 OPTISYS SLM 2100 5.2 Switching on the power DANGER! To avoid fatal injuries as well as destruction or damage of the device assure a correct installation before switching on the power. This includes: • The device is mechanically safe, mounting and power connection comply with the regulations. • Both doors need to be closed and locked. • The electrical operating data of the power supply comply with the requirements of the device.
.book Page 38 Wednesday, April 30, 2014 1:39 PM 5 OPERATION OPTISYS SLM 2100 5.3 Operating elements The operation elements consist of the four operation keys below the display.
.book Page 39 Wednesday, April 30, 2014 1:39 PM OPERATION 5 OPTISYS SLM 2100 INFORMATION! • Always push the keys at right angles to the front. Touching them from the side can cause incorrect operation! • After 5 minutes of inactivity in the menu mode, there is an automatic return to the measuring mode. In this case the device does not save previously changed data! 5.4 Measuring page After switching on the power, the display always shows the measuring page.
.book Page 40 Wednesday, April 30, 2014 1:39 PM 5 OPERATION OPTISYS SLM 2100 5.4.1 Navigating through the menus To switch from the measuring mode to the menu mode you have to execute the following steps: • Press > for at least 2.5 seconds (meanwhile the display shows the text "keep key pressed" and a time counting back). • After 2.5 seconds release the pressed key (you have to do this within the next 2.5 seconds during which the display shows the text "release key now" and a time counting back).
.book Page 41 Wednesday, April 30, 2014 1:39 PM OPERATION 5 OPTISYS SLM 2100 Display in submenu level Figure 5-4: Example for selection in submenu menu level 1 Indicates a possible status message in the status list (only if there are one or more messages) 2 Superior main menu 3 Menu code of selected menu item (see table of menu structure or function tables); up to two figures can appear right beside the character indicating the main menu (one if you are on the first submenu level and two on the second s
.book Page 42 Wednesday, April 30, 2014 1:39 PM 5 OPERATION OPTISYS SLM 2100 INFORMATION! Like the lowest menu level (i.e. the parameter or data level) some submenus also offer the possibility to make adjustments. The procedure is similar to the procedure on the parameter level. On the parameter level you see a list with one or more parameters or data. You can scroll through this list using ↑ or ↓: INFORMATION! In this list all parameters or data which have been changed (i.e.
.book Page 43 Wednesday, April 30, 2014 1:39 PM OPERATION 5 OPTISYS SLM 2100 Display with parameter or data ready to edit Figure 5-6: Example of parameter or data ready to edit 1 2 3 4 5 6 7 8 9 Indicates a possible status message in the status list (only if there are one or more messages) Superior submenu or subsequent submenu Menu code of selected parameter (see table of menu structure or function tables) Symbol denotes that there is a factory setting Symbol denotes that there is a permissible value
.book Page 44 Wednesday, April 30, 2014 1:39 PM 5 OPERATION OPTISYS SLM 2100 5.5 Menu mode structure INFORMATION! The following table just presents an overview. When programming the device, always consult the function tables additionally as they contain further information! Measuring Main menu mode Submenu Parameter 3 or 4 pages, scrolling with ↓ or ↑ > A1 language ^ A2 set clock A3 reset errors > ^ > 2.5 s A quick setup ^ ↓↑ > 2.5 s B test ^ ↓↑ > B1 simulation ^ B2 actual values B2.
.book Page 45 Wednesday, April 30, 2014 1:39 PM OPERATION 5 OPTISYS SLM 2100 5.6 Function tables 5.6.1 Menu A, quick setup A1, language Language selection (depends on the region for which the device has been ordered). Available languages: German, English. A2, set clock Manual setting of date and time. A3, reset errors This functions allows to reset all errors that are not reset automatically (e.g. power fail or counter overflow).
.book Page 46 Wednesday, April 30, 2014 1:39 PM 5 OPERATION OPTISYS SLM 2100 B3, logbooks Level Designation / function Settings / descriptions This menu groups several functions which allow to display the corresponding actual reading. B3.1 status log Log for information, errors and warnings. B4, information Level Designation / function Settings / descriptions This menu groups several other menus which contain device specific information. B4.1 Device serial no. B4.2 Mainboard B4.
.book Page 47 Wednesday, April 30, 2014 1:39 PM OPERATION 5 OPTISYS SLM 2100 C2, maintenance Level Designation / function Settings / descriptions This menu groups several other menus in which the maintenance functions can be set. C2.1 Calibration In this menu the concentration measurement can be calibrated. For further information refer to Calibration on page 55.
.book Page 48 Wednesday, April 30, 2014 1:39 PM 5 OPERATION OPTISYS SLM 2100 Level Designation / function C3.4.2 Invert signal C3.5 Control input 3 C3.5.1 Mode This function allows to define the different modes of the corresponding control input. Options: • off • maintenance C3.5.2 Invert signal This function allows to define how the control input is activated.
.book Page 49 Wednesday, April 30, 2014 1:39 PM OPERATION 5 OPTISYS SLM 2100 C4, extended setup Level Designation / function Settings / descriptions This submenu allows to set the initial parameters for commissioning of the meter. C4.1 Maximum depth Definition of the maximum depth. C4.2 Start position Definition of the start position. C4.3 Blind zone Definition of the blind zone. C4.4 Rake height Definition of the rake height. C4.5 Cleaning Settings of the corresponding relay. C4.5.
.book Page 50 Wednesday, April 30, 2014 1:39 PM 5 OPERATION OPTISYS SLM 2100 C5, device Level Designation / function Settings / descriptions This menu groups all functions that have no effect on the measurement or any output directly. C5.1 Device info Information about the device. C5.1.1 Device serial no. This menu shows the serial number of the device. C5.1.2 Mainboard C5.1.3 Sensor electronic C5.2 Display Groups several function for the fine adjustment of the display. C5.2.
.book Page 51 Wednesday, April 30, 2014 1:39 PM OPERATION 5 OPTISYS SLM 2100 5.7 Functions in detail 5.7.1 Definition of positions and zones Figure 5-7: Definition of zones and positions 1 Home position: this position marks the home (or zero point position) of the sensor. After any change of the settings and to calibrate the zero point for the level measurement, the sensor will always return to this position.
.book Page 52 Wednesday, April 30, 2014 1:39 PM 5 OPERATION OPTISYS SLM 2100 5.7.2 Measuring function In this menu point the measuring function can be selected. There are 4 different settings: 1. Level mode: mode In this mode the level of the sludge blanket and the fluff is monitored. Thus the meter measures the height of two distinctly different concentrations simultaneously.
.book Page 53 Wednesday, April 30, 2014 1:39 PM OPERATION 5 OPTISYS SLM 2100 5.7.4 Definition of fluff and blanket concentration In this menu the concentration of the fluff respectively the sludge blanket has to be set. In level mode these two concentrations are monitored. During the measurement cycle the Fluff concentration is tracked first, while the sensor is moving down. The factory settings of these values are: 1. Concentration fluff: 0.5 g/l 2. Concentration sludge blanket: 5 g/l 5.7.
.book Page 54 Wednesday, April 30, 2014 1:39 PM 5 OPERATION OPTISYS SLM 2100 DANGER! To avoid damages or destructions, use a limit switch in all applications where a rake or other moving devices may come in contact with the sensor or cable! The manufacturer does not assume any responsibility for damages caused by the absence or malfunction of the rake guard limit switch. 5.7.7 Timer / start signal If there is no danger for the sensor or cable to be caught by moving devices (i.e.
.book Page 55 Wednesday, April 30, 2014 1:39 PM OPERATION 5 OPTISYS SLM 2100 5.7.9 Calibration INFORMATION! Before leaving the factory the sensor is calibrated to offer an accurate measurement. Though it might be necessary to re-calibrate the sensor during the lifetime of the system. INFORMATION! Even after the installation, a recalibration might improve the accuracy. There are two forms of calibration.
.book Page 56 Wednesday, April 30, 2014 1:39 PM 5 OPERATION OPTISYS SLM 2100 Step 3: calibrate zero Press or until the submenu calibrate zero is highlighted. Press > to enter the chosen menu. actual value is shown. Press ^ to enter the chosen menu. calibrate zero is shown. Press or to select calibrate to calibrate or break to interrupt the calibration. Press ^ to enter the chosen menu. The calibration zero is done. • Press ^ several time until you reach the measuring mode again.
.book Page 57 Wednesday, April 30, 2014 1:39 PM OPERATION 5 OPTISYS SLM 2100 Step 3b: calibrate conc. without known reference value Start the calibration process with step 2: move sensor Press or until the submenu calibrate conc. is highlighted. Press > to enter the chosen menu. reference value is shown. Press ^ to skip this function. ADC value is shown. Press or to choose between stored value to use the stored value or actual value to use the actual value for calibration.
.book Page 58 Wednesday, April 30, 2014 1:39 PM 5 OPERATION OPTISYS SLM 2100 5.7.
.book Page 59 Wednesday, April 30, 2014 1:39 PM OPERATION 5 OPTISYS SLM 2100 5.8 Status messages and diagnostic information INFORMATION! Note the following information concerning the tables with the status messages in the following subsections: • The devices groups all error messages and maps them according to their category and importance; the following subsections and their tables represent the categories and the order of importance.
.book Page 60 Wednesday, April 30, 2014 1:39 PM 5 OPERATION OPTISYS SLM 2100 Explanation of the error categories Letter on display Error category according to: Description Action NE64 VDI/VDE 2650 and NE107 Bold "F F" Failure Device failure Whole device or single component(s) defective, replacement necessary, measured values are not valid; mA output ≤ 3.6 mA or set fault current (depending on the seriousness of the fault).
.book Page 61 Wednesday, April 30, 2014 1:39 PM OPERATION 5 OPTISYS SLM 2100 5.8.2 Error category "Application error" ( "F", not bold) INFORMATION! Regard that also in case of an error message of the category "F" the measured values are not valid! Messages on the display Description Actions F Door open The door of the cable drum compartment is open. Close the door; if closed check interlock switch; Error will be automatically reset, if problem was solved.
Service .book Page 62 Wednesday, April 30, 2014 1:39 PM 6 SERVICE OPTISYS SLM 2100 6.1 Manual initiation of a measurement For service or maintenance you can manually initiate a measurement: • Press ↓ on the Keypad for five seconds. i The meter starts with a measurement cycle as soon as possible: the sensor goes down to the defined maximum depth and returns to its start position, afterwards the meter works in the normal mode. 6.
.book Page 63 Wednesday, April 30, 2014 1:39 PM SERVICE 6 OPTISYS SLM 2100 Check of screws and nuts Check if all screws of the reeling unit as well as on the mounting frame are properly tight and secured. Sensor check in water Check the sensor by immersion in water; this should give a low measuring value or even a "0" on the converter display. By placing something solid (e.g. your finger) in the gap between the LED and the sensor you should get a measuring value of 3000...
.book Page 64 Wednesday, April 30, 2014 1:39 PM 6 SERVICE OPTISYS SLM 2100 Cleaning or replacement of air filter DANGER! If the air filter is blocked the meter can catch fire which could result in health and safety hazards or damage of the plant equipment. WARNING! Turn off the device during maintenance of the fan and fan filter. 1. Loosen the 2 nuts in the ventilation shaft. 2. Remove the ventilation shaft by pulling it up. 3. Remove the cover of the fan. 4. Clean or replace the air filter.
.book Page 65 Wednesday, April 30, 2014 1:39 PM SERVICE 6 OPTISYS SLM 2100 6.3 Spare parts availability The manufacturer adheres to the basic principle that functionally adequate spare parts for each device or each important accessory part will be kept available for a period of 3 years after delivery of the last production run for the device. This regulation only applies to spare parts which are subject to wear and tear under normal operating conditions. 6.
.book Page 66 Wednesday, April 30, 2014 1:39 PM 6 SERVICE OPTISYS SLM 2100 6.5.2 Form (for copying) to accompany a returned device Company: Address: Department: Name: Tel. no.: Fax no.: Manufacturer's order no. or serial no.: The device has been operated with the following medium: This medium is: water-hazardous toxic caustic flammable We checked that all cavities in the device are free from such substances. We have flushed out and neutralized all cavities in the device.
Technical data .book Page 67 Wednesday, April 30, 2014 1:39 PM TECHNICAL DATA 7 OPTISYS SLM 2100 7.1 Measuring principle The meter is using an optical sensor which travels through the media. Thus it can directly measure the suspended solids concentration at different heights. The measurement of the suspended solids content is based on the method of the transmission of light, which provides precise measurement results independent of the sludge colour.
.book Page 68 Wednesday, April 30, 2014 1:39 PM 7 TECHNICAL DATA OPTISYS SLM 2100 7.2 Technical data INFORMATION! • The following data is provided for general applications. If you require data that is more relevant to your specific application, please contact us or your local sales office. • Additional information (certificates, special tools, software,...) and complete product documentation can be downloaded free of charge from the website (Download Center).
.book Page 69 Wednesday, April 30, 2014 1:39 PM TECHNICAL DATA 7 OPTISYS SLM 2100 Measuring accuracy Maximum measuring error (concentration) 5% full scale Maximum measuring error (level) 1% full scale Operating conditions Altitude ≤ 2000 m / 6561.68 ft Temperatures and pressure Ambient temperature (in operation) -20...+50°C / -4...+122°F Process temperature (sensor) 0...+60°C / +32...+140°F Storage temperature 0...+60°C / +32...+140°F Humidity Max. 90% at 40°C / 104°F Max.
.book Page 70 Wednesday, April 30, 2014 1:39 PM 7 TECHNICAL DATA OPTISYS SLM 2100 Inputs and outputs General All in and outputs are electrically isolated from the power supply. Inputs Control inputs: Three active control inputs, not polarity sensitive U out = 8 VDC I out = -10 mA U low = <2 V min I typical at 2 V = -8.7 mA U high = >4 V max I typical at 4 V = -6.
.book Page 71 Wednesday, April 30, 2014 1:39 PM TECHNICAL DATA 7 OPTISYS SLM 2100 Approvals and certifications CE EMC Low voltage directive The device fulfils the statutory requirements of the EC directives. The manufacturer certifies that these requirements have been met by applying the CE marking.
.book Page 72 Wednesday, April 30, 2014 1:39 PM 7 TECHNICAL DATA OPTISYS SLM 2100 7.3 Dimensions Figure 7-2: Dimensions Dimensions [cm] Dimensions ["] Weight [kg] 72 a 460 18.11 b 420 16.54 c 188.55 7.1 d 512 20.16 e 81 3.19 f 260 10.24 g 550 21.65 h Ø 85 3.35 j Ø 40 1.57 k 231.5 9.11 www.krohne.com [lbs] 26.5 58.
Notes .book Page 73 Wednesday, April 30, 2014 1:39 PM NOTES 8 OPTISYS SLM 2100 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en www.krohne.
.book Page 74 Wednesday, April 30, 2014 1:39 PM 8 NOTES 74 OPTISYS SLM 2100 www.krohne.
.book Page 75 Wednesday, April 30, 2014 1:39 PM NOTES 8 OPTISYS SLM 2100 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en www.krohne.
.book Page 76 Wednesday, April 30, 2014 1:39 PM © KROHNE 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en - Subject to change without notice.