Owner manual

2 INSTALLATION
18
OPTIWAVE 5200 C/F
www.krohne.com 04/2013 - 4001905802 - AD ATEX OPTIWAVE 5200 R02 en
Compact and Remote (Field) versions - Metallic Horn and Wave Guide antennas
Equipment category 1/2D, 2D or EPL Da/Db, Db: Ex ia and Ex ia tb devices
Equipment category 3 D or EPL Dc: Ex ic devices
Flange
temperature
Ambient temperature
Metallic Horn (standard temperature)
Wave Guide
Metallic Horn (high-temperature)
Compact version Remote (Field)
version
Compact version Remote (Field)
version
[°C] [°F] [°C] [°F] [°C] [°F] [°C] [°F] [°C] [°F]
+80 +176 +80 +176 +80 +176 +80 +176 +80 +176
+90 +194 +75 +167 +75 +167 +78 +172 +79 +174
+100 +212 +71 +160 +70 +158 +77 +171 +77 +171
+110 +230 +66 +151 +65 +149 +75 +167 +76 +169
+120 +248 +62 +144 +60 +140 +74 +165 +75 +167
+130 +266 +57 +135 +55 +131 +72 +162 +73 +163
+140 +284 +53 +127 +51 +124 +71 +160 +72 +162
+150 +302 +48 +118 +46 +115 +69 +156 +71 +160
+160 +320 - - - - +67 +153 +69 +156
+170 +338 - - - - +66 +151 +68 +154
+180 +356 - - - - +64 +147 +67 +153
+190 +374 - - - - +63 +145 +65 +149
+200 +392 - - - - +61 +142 +64 +147
+210 +410 - - - - +60 +140 +63 +145
+220 +428 - - - - +58 +136 +61 +142
+230 +446 - - - - +56 +133 +60 +140
+240 +464 - - - - +55 +131 +59 +138
+250 +482 - - - - +53 +127 +57 +135