Instruction Manual

OPERATION 5
11
OPTIWAVE 7300 C
www.krohne.com01/2013 - 4002495601 - AD WHG OPTIWAVE 7300 R02 en
5.3.3 Device settings (DTM)
When you change the configuration of a WHG-approved device with the DTM and then store to
the device, you must confirm the change 2 times. Do the procedure that follows:
Store to the device (procedure for WHG-approved devices)
Change the device settings.
Click on the Store to device button.
i The DTM opens a "Save" window and tells the user to make sure it is necessary to store the
changes.
If it is necessary to store the changes, click on the Yes button.
i The DTM opens a WHG WARNING window.
If the device settings are correct, click on the Store to device button.
i The DTM opens a second WHG WARNING window.
If the device settings are correct, click on the Store to device button.
i The device stores the new data to the device. Some minutes are necessary to complete this
step. End of the procedure.
Figure 5-3: Store to device button