3.3 Installation procedure
Because a new ORP sensor needs to be calibrated before it is installed into its final measuring
location, it is important to follow the installation order:
1 Unpack the sensor.
2 Connect the sensor to the junction box or directly to the process control system.
3 Calibrate the sensor.
4 Install the sensor into its final measuring location.
The required steps are explained in the following sections.
3.4 Pre-installation requirements
Unpacking the sensor
Gently pull the protection cap from the sensor 1.
Lay the sensor on a soft mat/tissue 2.
Keep the protection cap in the original packaging 3.
WARNING!
During installation of the device make sure that you use ESD (electronic discharge) protection
equippment.
CAUTION!
Never touch or scratch the platinum electrode of the sensor.
Store the sensor in its original packaging in a dry, dust-free location. Keep it away from dirt. If
necessary, clean it as described on page 20
.