4 OPERATION
18
SMARTSENS ORP 1590
www.krohne.com 11/2013 - 4002547102 - MA SMARTSENS ORP 1590 R02 en
4.1.2 Calibration with HART
®
Handheld 475 FIELD COMMUNICATOR
Manual hold
Start the function calibration
calibrationcalibration
calibration in menu mode Quick Setup
Quick SetupQuick Setup
Quick Setup or Setup
SetupSetup
Setup.
Activate the function manual hold
manual holdmanual hold
manual hold to avoid an alarm.
Select Yes
YesYes
Yes.
Press Enter
EnterEnter
Enter to proceed.
Start calibration procedure
Set value of ORP solution; Default setting: 468 mV. Press the key >
>>
> to enter the submenu. Now
you can change the default setting.
Press Enter
EnterEnter
Enter and Ok
OkOk
Ok to proceed.
Dip the sensor into the ORP solution and wait till the value is stable.
Press Ok
OkOk
Ok to proceed.
i The measurement of ORP solution starts. After approx.10 seconds and a successful
calibration the following values appear:
Old ORP value in mV
ORP solution in mV
Electrode voltage in mV
New offset in mV
New ORP value in mV
Press Ok
OkOk
Ok to proceed.
Set the calibration date with DD-MM-YYYY. Press the key >
>>
> to enter the submenu. Now you can
change the calibration date .
Press Enter
EnterEnter
Enter and Ok
OkOk
Ok to proceed.
i The message Save values?
Save values?Save values?
Save values? appear.
Press yes
yesyes
yes to save the values.
Press Enter
EnterEnter
Enter to proceed.
Select no
nono
no to deactivate the function manual hold.
Press Enter
EnterEnter
Enter to leave the calibration menu.
A ORP calibration is necessary in regular intervals or when installing a new ORP sensor.
If the calibration was not successful, the message Warning!!!
Warning!!!Warning!!!
Warning!!! appears behind the value New
New New
New
offset
offsetoffset
offset.
Press Ok
OkOk
Ok to proceed.
Set the calibration date with DD-MM-YYYY. Press the key >
>>
> to enter the submenu. Now you can
change the calibration date .
Press Enter
EnterEnter
Enter and Ok
OkOk
Ok to proceed.
i The messages Warning!!! Offset out of range!
Warning!!! Offset out of range!Warning!!! Offset out of range!
Warning!!! Offset out of range! and Save values?
Save values?Save values?
Save values? appear.
Select betweenyes
yesyes
yes or no
nono
no.
CAUTION!
Make sure that the sensor tip is clean and dust-free. If necessary, clean the tip as described on
page 20
.
CAUTION!
Moisture inside the sensor connector must be avoided! Moisture may cause a shortcut and
deliver erratic readings!
If moisture has entered the connector dry it with air (e.g. hair blower).