Press Next
NextNext
Next to proceed.
Select no
nono
no to deactivate the function manual hold
manual holdmanual hold
manual hold.
Press Next
NextNext
Next to leave the calibration menu.
In a fully functional sensor, the optimal slope is -59 mV for each pH unit and the optimal zero
point is 0 mV at pH 7. The slope should at least have a value between -50...-65 mV per pH unit.
Re-calibrate the sensor if the slope does not approximate those limits. The pH sensor ages, the
slope gets flatter and the zero error increases.
4.1.2 Calibration with HART
®
Handheld 475 FIELD COMMUNICATOR
Manual hold
Start the function calibration
calibrationcalibration
calibration in menu mode Quick Setup
Quick SetupQuick Setup
Quick Setup or Setup
SetupSetup
Setup.
Activate the function manual hold
manual holdmanual hold
manual hold to avoid an alarm.
Select Yes
YesYes
Yes.
Press Enter
EnterEnter
Enter to proceed.
Temperature compensation
Set the temperature compensation
temperature compensationtemperature compensation
temperature compensation for calibration.
Select between automatic
automaticautomatic
automatic or manual
manualmanual
manual.
i If you have selected manual temperature compensation
manual temperature compensationmanual temperature compensation
manual temperature compensation, please enter the temperature of
the pH buffer solution. Default setting: 25°C / 77°F.
Press Enter
EnterEnter
Enter to proceed.
Start calibration procedure
Dip the sensor into the first pH buffer solution 1 and wait till the value is stable.
Press Ok
OkOk
Ok to proceed.
i The measurement of pH buffer solution 1
pH buffer solution 1pH buffer solution 1
pH buffer solution 1 starts and the values pH and temperature appear.
After approx. 10 seconds the following values appear:
pH value
temperature value
Status buffer detection Finished
Value indentified buffer
Press Ok
OkOk
Ok to confirm the identified pH buffer solution. If neccessary change the value of the pH
buffer solution manually.
Press Enter
EnterEnter
Enter to proceed.
i If the sensor does not recognise the pH buffer solution the messages
Identified buffer
Identified buffer Identified buffer
Identified buffer 0 pH
0 pH 0 pH
0 pH and Buffer not recognised
Buffer not recognisedBuffer not recognised
Buffer not recognised appear.
Press Ok
OkOk
Ok to proceed.
i Enter the value of the pH buffer solution manually.
CAUTION!
Never touch or scratch the pH sensitive glass tip of the sensor.
Make sure that the sensor tip is clean and dust-free. If necessary, clean the tip as described
on page 24
.
CAUTION!
Moisture inside the sensor connector must be avoided! Moisture may cause a shortcut and
deliver erratic readings!
If moisture has entered the connector dry it with air (e.g. hair blower).