User guide

4 OPERATION
20
SMARTSENS PH 8570
www.krohne.com 11/2013 - 4002544902 - MA SMARTSENS PH 8570 R02 en
Press Enter
EnterEnter
Enter to proceed with the calibration procedure.
Clean the sensor with demineralised water 2.
Dip the sensor into the second pH buffer solution 3 and wait till the value is stable.
Press Ok
OkOk
Ok to proceed.
i The measurement of pH buffer solution 2
pH buffer solution 2pH buffer solution 2
pH buffer solution 2 starts and the values pH and temperature appear.
After approx. 10 seconds the following values appear:
pH value
temperature value
Status buffer detection Finished
Value indentified buffer
Press Ok
OkOk
Ok to confirm the identified pH buffer solution. If neccessary change the value of the pH
buffer solution manually.
Press Enter
EnterEnter
Enter to proceed.
i If the sensor does not recognise the pH buffer solution the messages
Identified buffer
Identified buffer Identified buffer
Identified buffer 0 pH
0 pH 0 pH
0 pH and Buffer not recognised
Buffer not recognisedBuffer not recognised
Buffer not recognised appear.
Press Ok
OkOk
Ok to proceed.
i Enter the value of the pH buffer solution manually.
Press Enter
EnterEnter
Enter to proceed with the calibration procedure.
Clean the sensor with demineralised water 2.
i After a successful calibration the following values appear:
Buffer solution 1
Buffer solution 2
Offset in mV
Slope in mV/pH
Press Ok
OkOk
Ok to proceed.
Set the calibration date with MM-DD-YYYY. Press the key > to enter the submenu. Now you can
change the calibration date.
Press Enter
EnterEnter
Enter and Ok
OkOk
Ok to proceed.
i The message save the values?
save the values?save the values?
save the values? appears.
Select yes
yesyes
yes to save the values.
Press Enter
EnterEnter
Enter to proceed.
Figure 4-2: Calibration procedure
.book Page 20 Monday, November 18, 2013 4:44 PM