3.6.6 Cleaning of flow sensor
The flow sensor is highly resistant against dirt and the measurement will rarely be influenced by
anything. However, it is advisable to create a possiblity for cleaning just in front or behind the sensor.
3.6.7 Flange deviation
3.6.8 Mounting position
Figure 3-9: Option for cleaning of flow sensor
1 Opening for cleaning
CAUTION!
Max. permissible deviation of pipe flange faces: L
max
- L
min
0.5 mm / 0.02"
Figure 3-10: Flange deviation
1 L
max
2 L
min
CAUTION!
Only install the flow sensor in the shown position to keep the electrodes under water. Limit the
rotation to
±
2
°
to maintain the accuracy.
Figure 3-11: Mounting position