User Manual
6 TECHNICAL DATA
36
TIDALFLUX 2300 F
www.krohne.com 06/2013 - 4002490202 MA TIDALFLUX 2300 F R02 en
Installation condtitions
Installation For detailed information see chapter "Installation".
Flow direction Forward and reverse.
Arrow on flow sensor indicates positive flow direction.
Inlet run ≥ 5 DN (without disturbing flow, after a single 90° bend)
≥ 10 DN (after a double bend 2x 90°)
≥ 10 DN (behind a control valve)
Outlet run ≥ 3 DN
Dimensions and weights For detailed information see chapter "Dimensions and weights".
Materials
Sensor housing Standard: sheet steel
Other materials on request
Measuring tube Austenitic stainless steel
Flange Standard: carbon steel, polyurethane coated
Other materials on request.
Liner Polyurethane
Connection box IP 67: polyurethane coated die-cast aluminium
IP 68: Stainless steel
Measuring electrodes
Hastelloy
®
C
Grounding rings Stainless steel
Tailor made to innerdiameter of connecting pipeline.
Necessary if innerside of connecting pipeline isn't electrically
conductive.
Process connections
Flange
FlangeFlange
Flange
EN 1092-1 DN200...1600 in PN 6...40 (others on request)
ASME 8...64" in 150...300 lb RF (others on request)
JIS DN200...1600 in JIS 10...20 K (others on request)
Design of gasket surface RF (others on request)