Instruction Manual

6 OPERATION
40
UFM 530 HT
www.krohne.com 12/2013 - 4002393201- HB UFM 530 HT R02 en
3.01.11 MIN VOS Minimum velocity of sound (VOS). Value used for I0% or P
0%
when
function "VOS" selected in function 3.04.01 or 3.05.01.
Units: m/s or feet/s,
range: 0...4999 m/s (0...16.400 feet/s).
3.01.12 MAX VOS Maximum velocity of sound.Value used for I
100%
or P
100%
when
function "VOS" selected in function 3.04.01 or 3.05.01. Note:
maximum value must be greater than minimum value!
Units: m/s or feet/s,
range: 1...4999 m/s (0...16.400 feet/s).
3.02.00, VERSION
3.02.01 FUNCTION Function of converter. This is factory pre-set and can only be changed
to standard from any setting:
1) STANDARD.
2) CORR T: temperature correction of the measured flow via analog
input 1.
3) CORR T + P: temperature and pressure correction of the measured
flow via analog input 1 (pressure) and analog input 2 (temperature).
4) HEAT: reserved for measurement of heat power and heat energy
totalisation.
5) BATCH: batch volume.
6) MODIS: Cannot be altered.
3.02.02 INP1 4 mA 4 mA reference for analog input 1, 4 mA temperature reference.
Unit: °Cor°F, range: -50...+150°C / -58...302°F.
3.02.03 INP1 20 mA 20 mA reference for analog input 1, 20 mA temperature reference.
Unit: °Cor°F, range: -50...+150°C / -58...302°F.
3.02.04 INP2 4 mA 4 mA reference for analog input 2, 4 mA temperature reference.
Unit: °Cor°F, range: -50...+150°C / -58...302°F.
3.02.05 INP2 20 mA 20 mA reference for analog input 2, 20 mA temperature reference.
Unit: °Cor°F, range: -50...+150°C / -58...302°F.
3.02.06 INP2 4 mA 4 mA reference for analog input 2, 4 mA pressure reference.
Unit: bar(a) or psi(a), range: 0...100 bar(a).
3.02.07 INP2 20 mA 20 mA reference for analog input 2, 20 mA pressure reference.
Unit: bar(a) or psi(a), range: 0...100 bar(a).
3.02.08 K0 Product constant K0.
Range: 10
-9
...10
9
.
3.02.09 K1 Product constant K1.
Range: 10
-9
...10
9
.
3.02.10 K2 Product constant K2.
Range: 10
-9
...10
9
.
3.02.11 DENSITY 15 Product density at T = 15°C/ 59°F.
Range: 500...2000 kg/m
3
.
3.02.12 FULL SCALE Heat measurement. Full scale setting and unit for heat power.
Available units: GJ/s, GJ/hr, MJ/s, MJ/hr, GCal/s, GCal/hr, MCal/s,
MCal/hr.
No. of menu Display text Setting/description and functions