Instruction Manual
OPERATION 6
41
UFM 530 HT
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
3.02.13 BATCH VOL Batch volume total size and units.
Availabe units: m
3
, Liter, US-Gallon, Barrel or user definable unit.
Range: 0.025...100000 m
3
.
3.03.00, DISPLAY.
3.03.01 DISP FLOW Display of flow. Available options:
1) RATE: full-scale units (flow is shown with the unit as set in function
3.01.01).
2) PERCENT: flow is shown as a percentage (0...100%) of the full
scale value.
3) NO DISPLAY: no flow is shown.
3.03.02 FUNCT TOT Function of totalizer. Two totalizers (counters) are available. The
totalizer values are incremented and stored once a second. The
following settings are available:
1) ACT FLOW: actual flow units, used for counting the total volume in
the totalizer (for each direction one totalizer is available, the sum of
the two can also be displayed).
2) CORR FLOW: the corrected flow is used for counting the total
volume in the totalizers.
3) POS BOTH: both the actual and the corrected flow are used for
counting the total volume in the totalizers. Note: both are only
counted in the positive direction!
3.03.03 DISP TOTAL Display of totalizer. A totalizer can be selected here for displaying:
1) TOTAL OFF: both totalizers are switched off, counting stops.
2) FORWARD: forward volume units.
3) REVERSE: reverse volume units.
4) BOTH: both totalizers, alternating.
5) SUM: sum of both totalizers.
6) BOTH + SUM: both totalizers und their sums show alternating.
7) NO DISPLAY: totalizers are not displayed but totalizers count.
3.03.04 TOTAL VOL Unit for volume totalizer. Note: the maximal value of the totalizer is
99999999 x 10 m
3
and will roll over to 0 at overrun!
Available units: X10 m
3
, US-Gallon, m
3
, barrel, liter.
3.03.05 TOTAL ENER Unit for the heat energy totalizer. Note: the maximal value of the
totalizer is 99999999 x 10 GJ and will roll over to 0 at overrun!
Available units: X10 GJ, GJ, MJ, GCal., MCal.
3.03.06 VOS Unit for velocity of sound. Available options:
1) NO DISPLAY: no display of the velocity of sound.
2) m/s, feet/s.
3.03.07 CYCL DISP If more than one measured value are to be displayed (e.g. flow rate
and totalizer), each value can be selected manually by pressing the
↑ key, or the values can be alternately displayed each five seconds by
turning the cyclic display function on.
3.03.08 ERROR MSG Display error messages
No. of menu Display text Setting/description and functions