Parameter error menu
3.09.00, COMMUNICATION
3.09.01 PROTOCOL Communication protocol:
1) OFF: no communication.
2) HART: HART
®
(standard available).
3) PROFIB PA: PROFIBUS PA (optional available).
3.09.02 HART ADDR
HART
®
address.
Range: 00...16.
3.09.03 PP/FF ADDR PROFIBUS PA address.
Range: 000...126.
INFORMATION!
In the next part, multiple error messages are shown. Details about the specific functions are
described earlier in the "Installation menu".
No. of menu Display text Setting/description and functions
4.01.00, FLOW VELOC
Volume flow velocity (v) value incorrect. The flow speed is calculated from the full scale volume flow and
the meter size. Note: ensure condition 0.5 m/s ≤ v ≤ 20 m/s (1.64 feet/s ≤ 65.62 feet/s) is met!
4.01.01 FULL SCALE Full-scale value for 100% volume flow rate.
4.01.02 METER SIZE Meter size.
4.02.00, CURR OUTP (Current output)
Current output range incorrect. Setting for 100% is compared with setting for 0%. Note: ensure condition
100 pct-0 pct ≥ 4mA is met!
4.02.01 RANGE Range of current output.
4.02.02 0pct Current value for 0% scale.
4.02.03 100 pct Current value for 100% scale.
4.03.00, LF CUTOFF (Low-flow cut-off)
Low-flow cut-off range incorrect. If low flow cut-off is set to "on", the value for CUTOFF-OFF is compared
with the value of CUTOFF-ON on. Note: ensure condition CUTOFF-OFF – CUTOFF-ON ≥ 1% is met!
4.03.01 LF CUTOFF Low-flow cut-off.
4.03.02 CUTOFF ON Cutoff "on" value.
4.03.03 CUTOFF OFF Cutoff "off" value.
4.04.00, ENERGY
Full scale value for heat energy rate (E) incorrect. The fullscale value is compared with the maximum
value that can be measured and should meet the condition E
max
<E
fullscale
<E
max
/1000. Note: the
maximum value that can be measured is at maximum flow and 200°C temperature difference!
4.04.01 HEAT FS Full-scale value for 100% heat energy rate.
4.05.00, PULSE/VOS (Pulse/Velocity of sound)
Unit of pulse output for velocity of sound function incorrect. Note: ensure that "PULSE RATE" is selected
for "VOS"!